13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

2.10 Dual Channel Measurements<br />

Dual channel measurements are available only when using the Model 8502A power meter. The dual<br />

channel capability of the Model 8502A provides two very useful functions. These are:<br />

• Display of power measured by two channels simultaneously.<br />

• Display of the power ratio between two channels.<br />

Applications for these two functions are many and varied. Dual channel measurements allow two points<br />

in a system to be measured with one power meter. The PPM can simultaneously measure both channels in<br />

either the CW or <strong>Peak</strong> <strong>Power</strong> Modes.<br />

The Ratio function allows relative measurements to be made in either CW or <strong>Peak</strong>. It is also possible to<br />

ratio <strong>Peak</strong> to CW or CW to <strong>Peak</strong>. These functions are useful for gain and loss measurements under CW<br />

or <strong>Peak</strong> conditions (or a combination of the two).<br />

2.10.1 CW, CW Measurements<br />

The following procedure is used for making dual channel CW measurements:<br />

1. After power-on, press [CW] to access the CW Mode. Press [MEM] (3) [F1] for the PPM to use<br />

its default values, if desired. Wait approximately 30 minutes for the PPM to warm up.<br />

2. Self-Calibrate both channels. Press [MENU] (1) [F3] [F1].<br />

3. Follow the instructions displayed by the PPM to connect the appropriate detector to the<br />

Calibrator when prompted.<br />

4. Connect the detectors to the device(s) or system to be tested.<br />

5. The PPM will now be displaying the CW power for channel A. To display the CW power for<br />

both channels, press [B].<br />

6. Select the frequency of operation or cal factor as necessary for detector response correction by<br />

pressing [FREQ].<br />

7. Enter any required display offsets.<br />

8. To return to single channel operation, use the Detector Select A and B keys as required.<br />

2.10.2 <strong>Peak</strong>, <strong>Peak</strong> Measurements: Method 1<br />

1. Follow the procedure given in Section 2.10.1, except press [PEAK] in Step 1 instead of CW.<br />

2. If the PPM is triggering, it will display the peak power in both channels. The following default<br />

settings are in effect if the PPM has been initialized.<br />

a. Internal Trigger from channel A set at -10 dBm (i.e. pulse at channel A causes trigger if<br />

greater than -10 dBm)<br />

b. Frequency = 1.00 GHz<br />

c. Chan A Cursor Delay = 1.0000 µs<br />

d. Chan B Cursor Delay = 1.0000 µs<br />

e. Spin knob active on channel A Cursor Delay as indicated by ∆ next to CSR DLY<br />

3. If the PPM’s New Data and Ready lights are flashing, a trigger is occurring and measuring<br />

should be possible. Try adjusting the channel A Cursor Delay by turning the spin knob until the<br />

CSR DLY reads 0.0000 ns. The Delay can then be increased slowly until the reading is being<br />

taken on the desired part of the pulse.<br />

2-32 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!