<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

4. Place the system into the <strong>Peak</strong> Mode by pressing [PEAK].<br />

5. The FREQ = XXXX portion of the PPM display must read 5.000 GHz ±0.020 GHz.<br />

D. RF Blanking Output.<br />

1. Connect a coaxial cable from the Scope to the RF BLANKING output on the PPM.<br />

2. Disconnect the Detector from the Signal Generator.<br />

3. The Scope should display a dc level of +5 V.<br />

4. Enter the Autozero function by pressing [MENU] [F3] [F2] [UNITS].<br />

5. During the Autozero cycle the Scope display must drop to near 0 V, and then return back to +5 V.<br />

5.3.11 Volume Adjust<br />

1. Adjust the VOLUME ADJ on the back of the PPM fully ccw.<br />

2. Press several keys. Very little (if any) sound (clicks) should be heard from the speaker when the<br />

keys are pressed.<br />

3. Adjust the VOLUME ADJ fully clockwise.<br />

4. Press several keys. The clicks should be very easy to hear when the keys are pressed.<br />

The volume of the beeps used to indicate success or failure of zeroing, calibration, and self-test will be<br />

set to the same level as the adjustment for the clicks.<br />

5-20 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009