13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

2.7 Self-Calibration and Auto-Zeroing<br />

The purpose of the PPM’s Self-Calibration function is to ensure that measurement accuracy will be<br />

maintained despite changes in detectors, changes in detector diode temperature, or other changes that<br />

might occur in the PPM’s analog circuits.<br />

During the time the instrument is performing its Self-Calibration routine, the RF output of the 1 GHz<br />

calibrator is accurately stepped from +20 to -30 dBm. Each step is measured by an internal standard and,<br />

at each point, the calibrator output power as measured by the RF detector is compared to the standard.<br />

The power readings are then digitally adjusted by the PPM microprocessor and stored in memory. For<br />

power levels below -30 dBm, the system relies on the Auto-Zero function as well as the inherent<br />

repeatability of the detector voltage output versus input power at these levels. (Refer to Section 4.2.7 for<br />

a complete description of the automatic calibration system.)<br />

To make any type of power measurements with the PPM it will be necessary to use the appropriate<br />

calibration and zeroing procedures to ensure the highest accuracy.<br />

After the instrument has warmed up for 30 minutes and the mode of operation has been selected, a<br />

detector self-calibration should be performed. In addition, if the ambient room temperature changes more<br />

than 5 °C from the temperature at which the system was last calibrated, self-calibration should also be<br />

done. If there has been a change in the temperature of the detector of more than 5 °C since it was last<br />

calibrated, the difference will be shown.<br />

2.7.1 Self-Calibration Procedure<br />

Connect the channel A detector to the 1 GHz Calibrator output. Press [MENU] [F3] to call up the Cal or<br />

Zero Detectors menu. The menu offers the following choices:<br />

• F1 to Calibrate the Detectors<br />

• F2 to Auto-Zero the Detectors<br />

• F3 to Return to Data Display<br />

Select F1 for the Detector Self-Calibration.<br />

☛ NOTE: The CLEAR key as well as F3, can also be used to return to data display.<br />

The display will show the last date and time the detector was calibrated. To proceed with the calibration,<br />

press any Units key. The Self-Cal process takes about 50 seconds, during which time the statement<br />

A (or B) CALIBRATION IN PROGRESS<br />

will be displayed.<br />

If a Model 8501A is in use, press [CLEAR] to return the routine to the data display (the last mode of<br />

operation). If the Model 8502A is being used, press [CLEAR] to bypass the channel A calibration and go<br />

to the channel B calibration.<br />

After successful completion of the Self-Cal function, the instrument will go back to displaying data in the<br />

last mode of operation. With the Model 8502A, either bypassing or successfully completing the channel<br />

A calibration will take the routine to the channel B calibration.<br />

Channel B calibration is done in the same manner as the channel A calibration. The date and time of the<br />

last B channel calibration will be presented, and the option will be given to either press any Units key to<br />

start the cal procedure or press [CLEAR] to return to the data display.<br />

2-14 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!