13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

RF Detectors<br />

detector element, make sure that the center conductor is centered in the housing and then reinsert<br />

the element.<br />

CAUTION<br />

Do not twist the element as you insert it. Doing so could damage the detector<br />

element pin or the center conductor of the detector housing assembly.<br />

4. Remove and discard the old PROM. Use the instructions furnished with the PROM Programming<br />

Accessory Kit to program the new PROM.<br />

5. Insert the new PROM into the PROM PC board so that pin #1 is in the position marked on the<br />

PC board (see Figure D-4).<br />

D.5.3<br />

Reassembly of Detectors<br />

Follow these instructions to reassemble the detector (see Figure D-4).<br />

CAUTION<br />

Do not twist the detector element while inserting it. This can damage the detector<br />

element pin or the detector housing assembly center conductor.<br />

1. Make sure that the detector element leads are straight. Screw the detector housing assembly into<br />

the cap assembly. Be very careful not to damage the leads of the element.<br />

2. Insert the element leads into the proper holes in the Pre-Amp board as shown in Detail B of<br />

Figure D-4. The lead from the center pin goes to hole 1, and the other lead goes to hole 2. Solder<br />

the leads in place.<br />

3. After the element has been installed, it may be necessary to readjust the pulse response of the<br />

amplifier due to a possible difference in the video resistance of the new element diodes in<br />

relationship to the old element diodes.<br />

To adjust the pulse response, connect the detector to a pulsed microwave source of about 0 dBm. The<br />

pulse must be very clean as this procedure assumes a perfect input pulse. The rise time of the pulse must<br />

be less than 6 ns, and it should have no overshoot or ringing. If the shape of the pulse is not perfect but<br />

is accurately known, this will do.<br />

Set the PPM in the Graph mode with the Internal Trigger at -10 dBm. Autoscale the display by pressing<br />

[GRAPH] and set the window to a width of 0.1 µs. If the pulse shape is wrong, adjust C3 on the<br />

Pre-Amp board in small increments and wait for the trace on the screen to redraw. After finding an<br />

optimum setting for C3, adjust R9 until the best shape is obtained. Do not move either control very far<br />

from the original setting. The end result for this procedure is to set the maximum rise time consistent<br />

with no overshoot or ringing. If the rise time cannot be set below 15 ns, a small overshoot of less than<br />

5% is acceptable.<br />

4. Slide the sleeve onto the cap assembly. Replace the sleeve holder plate. Insert and tighten screws<br />

(1) and (2), and return the detector to service.<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 D-9<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!