Operation<br />
2.6 Mode Selection<br />
Follow these procedures to select the desired operating mode. The modes are CW, PEAK and GRAPH.<br />
Throughout this chapter, you will be instructed to press the MENU and MEM keys. The instruction may<br />
be followed by a number in parenthesis. This number corresponds to the menu number and indicates the<br />
number of times you need to press the key to reach the necessary menu.<br />
2.6.1 CW Mode<br />
If in the <strong>Peak</strong> or Graph mode, press [CW] to access the CW Mode.<br />
If a menu is being displayed and you want to exit it and go to the CW Mode press [CLEAR] [CW].<br />
If the 8502A is in the Ratio mode, selecting CW will cancel RATIO and the 8502A will display the CW<br />
power of channel A.<br />
In certain instances the option will be given to select the CW Mode if there is some reason that another<br />
mode is not operating properly. For example, if a trigger had not been provided as required in the <strong>Peak</strong><br />
Mode. The purpose is to allow an escape from erroneous modes of operation.<br />
Very few adjustments or menu selections are required to set up the unit for CW measurements. The two<br />
major settings that must be specified are defined as follows:<br />
Frequency<br />
Press [FREQ] to select the operating frequency at which the power will be measured. Normally, the PPM<br />
automatically adjusts for detector frequency response variations by using the factory measured frequency<br />
response data stored in the PROM in the detector housing. Other options are available as detailed in<br />
PROM Frequency Correction Section 2.8.2. Not compensating for the frequency response of the detector<br />
could result in an error in the power measurement reading.<br />
CW Averaging<br />
This setting selects how many samples of the signal being measured will be taken and averaged to reduce<br />
noise effects at low power levels. To adjust the averaging number, press [Menu] (2) [F1]. A number<br />
between 1 and 999 averages must then be entered. The number of averages presently in effect is not<br />
indicated until the averaging menu has been entered.<br />
Low <strong>Power</strong> Level Measurements<br />
When the PPM is used in the CW mode for power levels below -30 dBm, its dynamic range is extended<br />
to below -40 dBm. Under low power testing conditions it is necessary to observe the correct auto-zeroing<br />
procedures to obtain the highest measurement accuracy. Zeroing should be done more often in<br />
environments where temperature changes are more drastic. See the Auto-Zero procedures in Section 2.7.3.<br />
2.6.2 <strong>Peak</strong> Mode<br />
While you are in the CW or Graph Mode, press [PEAK] to access the <strong>Peak</strong> mode.<br />
If a menu is displayed and you want to exit and go to the <strong>Peak</strong> mode, press [CLEAR] [PEAK].<br />
After this, follow the trigger setting guidelines given below. Lack of a trigger can freeze the display. To<br />
exit either of these situations either the trigger can be set, or the CW Mode can again be accessed by<br />
pressing [CW].<br />
To make measurements in the <strong>Peak</strong> or Graph Mode, first ensure that the system is receiving a trigger<br />
signal that will drive its timebase and crystal-based delay generator. The delay generator accurately<br />
places a sampling window at the desired time on the pulse to be measured. The different triggering<br />
methods are detailed below.<br />