Testing and Calibration<br />

5.3.7 A3 GPIB/CAL Control Board<br />

CAUTION<br />

The maximum output power of the Calibrator is over 100 mW. Be sure it is working<br />

properly before connecting the power meter.<br />

A. Calibrator Frequency Check.<br />

1. Connect the Frequency Counter to the PPM Calibrator output.<br />

2. Select Calibrator TEST MODE by pressing [MENU] [F3] [F1] [7].<br />

3. The frequency must be 1 GHz ±50 MHz.<br />

4. Disconnect the Frequency Counter from the Calibrator output.<br />

5. Press [CLEAR].<br />

B. Calibrator Output Adjust.<br />

1. Zero the <strong>Power</strong> <strong>Meter</strong>.<br />

2. Connect the <strong>Power</strong> <strong>Meter</strong> detector to the Calibrator output on the PPM and press [7].<br />

3. Adjust A3R16 (CALIB SET) for 0 dBm.<br />

a. Press any numeric key. If the power meter indication is negative (e.g. -0.2 dBm), adjust<br />

A3R16 clockwise to approximately twice the error.<br />

b. Press any of the numeric keys to take a reading.<br />

c. Repeat Steps a and b until the power meter reads 0.00 dBm.<br />

4. Disconnect the detector from the PPM.<br />

5. Press [CLEAR].<br />

C. <strong>Power</strong> <strong>Meter</strong> Calibrate.<br />

1. Connect the Detector to the Calibrator output (use channel A if the unit is an 8502A).<br />

2. Press [UNITS].<br />

3. Wait for the PPM to calibrate itself (about 1 minute).<br />

4. Disconnect the detector from the Calibrator output.<br />

5. If the unit is an 8502A, repeat Steps 1 through 4 for channel B.<br />

5.3.8 A4 CPU Board<br />

A. Time Oscillator Adjust.<br />

1. Connect the Frequency Counter/Timer common lead to A4TP3 (COMMON).<br />

2. Connect the Frequency Counter/Timer test lead to A4TP6.<br />

3. Adjust A4C47 for 30.517578 µs ±0.000100 µs.<br />

4. Disconnect the Frequency Counter/Timer test leads.<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 5-15<br />

Superceded by Revision D, March 2009