Manual - 8500A Series Peak Power Meter - Giga-tronics
Manual - 8500A Series Peak Power Meter - Giga-tronics
Manual - 8500A Series Peak Power Meter - Giga-tronics
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
6<br />
Maintenance<br />
6.1 Introduction<br />
This chapter defines maintenance practices and troubleshooting procedures required for fault isolation<br />
down to PC board level.<br />
Problems can occur that might be produced by equipment or components peripheral to the 8501A/8502A<br />
units. Preliminary checks should be made to be sure that a malfunction of external equipment or<br />
components is not causing what appears to be a problem within the instrument.<br />
6.2 Periodic Maintenance<br />
The following maintenance procedures should be performed once each year unless the meter is operated<br />
in an extremely dirty or chemically contaminated environment. More frequent periodic maintenance may<br />
then required.<br />
1. Blow out all accumulated dust and dirt with forced air under moderate pressure. Also, the air<br />
filter for the cooling fan on the rear panel of the PPM should be checked every 3 months for<br />
accumulated dust. A dirty filter can restrict air flow into the unit and cause the instrument to<br />
overheat, which may cause the thermal sensor to shut the instrument off in the middle of a test<br />
routine. If dirty, the filter should be removed and cleaned with a mild detergent and water<br />
solution. Replacement filters can be ordered from the factory.<br />
2. Remove the filter from its holder by snapping the retainer out of the holder. Do not remove the<br />
screws from the filter holder.<br />
3. Inspect the unit for loose wires or damaged components. Check that the PC boards are properly<br />
seated in their receptacles, and that all wire connectors are properly attached to their PC board<br />
pins.<br />
4. The front panel and housing of the unit can be cleaned with a cloth dampened in a mild<br />
detergent. Do not use abrasive cleaners, scouring powders, or any harsh chemicals. Wipe the soap<br />
residue off with a clean, damp cloth, then dry with a clean dry cloth.<br />
5. If calibration is required, refer to Chapter 5 of this manual for procedures.<br />
6. Make a performance verification test in accordance with the procedures in Chapter 5. If the<br />
performance is within the required specifications, no further service is necessary.<br />
<strong>Manual</strong> No. 20790, Rev C, November 1998 6-1<br />
Superceded by Revision D, March 2009