A - Appendix A<br />

Configuring and Using the 3D Simulation Tool EVOSim<br />

To install it on another PC, copy the directory c:\Program<br />

Files\Tecan\<strong>EVOware</strong>\Simulation with all of the files and sub-directories to the<br />

other PC. In this case, EVOSim and EVOSim Splitter are started by doubleclicking<br />

on the respective .exe files in the chosen target directory.<br />

Please note the following points:<br />

• EVOsim can be used to simulate a <strong>Freedom</strong> EVO pipetting instrument.<br />

• <strong>Freedom</strong> <strong>EVOware</strong> does not require a hardlock (dongle) when EVOsim is<br />

simulating the pipetting instrument.<br />

• You must start <strong>Freedom</strong> <strong>EVOware</strong> in real mode to use EVOSim (firmware<br />

commands such as movement commands are not sent to the RS232 port in<br />

<strong>Freedom</strong> <strong>EVOware</strong>’s offline mode).<br />

• <strong>Freedom</strong> <strong>EVOware</strong> will run in offline mode if it can not detect either a pipetting<br />

instrument or EVOSim, which simulates the data communication with a<br />

pipetting instrument. If you are using <strong>Freedom</strong> <strong>EVOware</strong> with EVOSim, this<br />

normally indicates that there is a problem with the EVOSim configuration or<br />

the data connection(s). Note: the <strong>Freedom</strong> <strong>EVOware</strong> operation mode is<br />

shown in the Status Bar (see 6.7.3 “Status Bar”, 6-20).<br />

• EVOSim does not currently simulate the following optional devices: Te-Shake,<br />

Te-MagS, Te-Stack, Te-Stack Te-MO and Te-VacS. If you want to use<br />

EVOSim, you must first deactivate these devices in the <strong>Freedom</strong> <strong>EVOware</strong><br />

Configuration Tool (see 8.4.1.1 “Device Information Tab”, 8-21). If you don’t<br />

do this, <strong>Freedom</strong> <strong>EVOware</strong> will report errors when it tries to communicate with<br />

these devices.<br />

• EVOSim does not support the Pressure Monitored Pipetting (PMP) hardware<br />

option. If PMP is activated in your <strong>Freedom</strong> <strong>EVOware</strong> configuration, <strong>Freedom</strong><br />

<strong>EVOware</strong> will display an error message when it starts and PMP will be<br />

disabled for the current <strong>Freedom</strong> <strong>EVOware</strong> session.<br />

• When using EVOSim, you must always start it before you start <strong>Freedom</strong><br />

<strong>EVOware</strong>. If you also use EVOSim Splitter, start EVOSim Splitter first, then<br />

EVOSim and finally <strong>Freedom</strong> <strong>EVOware</strong>.<br />

• If you have been using EVOSim, you must always exit <strong>Freedom</strong> <strong>EVOware</strong><br />

before you exit EVOSim (<strong>Freedom</strong> <strong>EVOware</strong> sends firmware commands<br />

during the shut-down procedure and expects RS232 communications to<br />

remain functional during this time).<br />

• EVOSim does not report pipetting errors to <strong>Freedom</strong> <strong>EVOware</strong> (such as DITI<br />

already mounted). The only errors it reports are incorrect positioning<br />

commands (“movement out of range”). Out of range movements are also<br />

indicated by displaying the respective arm (LiHa, RoMa, PnP) in red. Note:<br />

For mechanical reasons, <strong>Freedom</strong> <strong>EVOware</strong> purposely sends movement<br />

commands which are out of range during pipetting instrument initialization.<br />

• If you want to run two copies of EVOSim on the same PC (not normally<br />

meaningful), first make a duplicate copy of the directory c:\Program<br />

Files\Tecan\<strong>EVOware</strong>\Simulation with all of the files and sub-directories. Do<br />

not start two copies of EVOSim from the same directory (conflicts are caused<br />

by the shared configuration file).<br />

A.14.1 Communication Between EVOSim and <strong>Freedom</strong> <strong>EVOware</strong><br />

Depending on requirements, the simulation tool can communicate with <strong>Freedom</strong><br />

<strong>EVOware</strong> in several different ways:<br />

A): EVOSim runs on the same PC as <strong>Freedom</strong> <strong>EVOware</strong>. The PC must have two<br />

unused RS232 ports. Connect the two ports using a null-modem cable. Configure<br />

<strong>Freedom</strong> <strong>EVOware</strong> to use one of the ports and configure EVOsim to use the other<br />

port.<br />

A - 104 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>