12.01.2014 Views

Software Manual Freedom EVOware Standard 2.3 Freedom ...

Software Manual Freedom EVOware Standard 2.3 Freedom ...

Software Manual Freedom EVOware Standard 2.3 Freedom ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

15 - Script Commands<br />

Periodic Wash Command<br />

Volume in Cleaner<br />

Specify the volume of system liquid which should be used to wash the outside<br />

of the tips (range: 0.1 - 100 ml). Washing takes place with the tips lowered into<br />

the cleaner of the specified wash station (tip height = Z-max). The wash cycle<br />

is skipped if you are flushing DITI adapters and Volume in Cleaner is ignored<br />

in this case.<br />

Use Waste<br />

Specify the worktable position for the waste you want to use by clicking on it in<br />

the Worktable Editor. You must first put a wash station with waste unit in the<br />

Worktable Editor at the required position. Alternatively, choose the required<br />

waste position from the pull down list. The list shows all waste units which are<br />

currently positioned in the Worktable Editor.<br />

The gray (protected) fields show the worktable coordinates of the selected<br />

waste.<br />

Use Cleaner<br />

Specify the worktable position for the cleaner you want to use by clicking on it<br />

in the Worktable Editor. You must first put a wash station with cleaner unit in<br />

the Worktable Editor at the required position. Alternatively, choose the<br />

required cleaner position from the pull down list. The list shows all cleaner<br />

units which are currently positioned in the Worktable Editor.<br />

The gray (protected) fields show the worktable coordinates of the selected<br />

cleaner.<br />

Choose a shallow cleaner if you only need to clean the ends of the tips.<br />

Choose a deep cleaner if there is a possibility of contamination along the shaft<br />

of the tip. The deep cleaner requires a larger volume of system liquid and<br />

cleaning takes somewhat longer.<br />

The wash cycle is skipped if you are flushing DITI adapters and Use Cleaner<br />

is ignored in this case.<br />

Use Fast-Wash Module<br />

If your instrument is fitted with FWO (Fast Wash Option) or MPO (Monitored<br />

Pump Option), check this checkbox to use the Fast Wash module instead of<br />

the dilutors to wash the tips.<br />

Use Pinch Valve<br />

If your instrument is fitted with the Low Volume Option, check this checkbox to<br />

activate the Low Volume Option after washing the tips.<br />

Period<br />

Specify the number of DITI replacement actions after which <strong>Freedom</strong><br />

<strong>EVOware</strong> should flush the DITI adapters.<br />

Delay in Waste<br />

To reduce the pressure inside the tubing, you can instruct the instrument to<br />

keep the valves open somewhat longer after flushing before moving the LiHa<br />

away from the waste (range: 0 - 1000 ms).<br />

Delay in Cleaner<br />

To reduce the pressure inside the tubing, you can instruct the instrument to<br />

keep the valves open somewhat longer after washing before moving the LiHa<br />

away from the cleaner (range: 0 - 1000 ms). The wash cycle is skipped if you<br />

are flushing DITI adapters and Delay in Cleaner is ignored in this case.<br />

Retract with<br />

Specify the speed with which you want to retract the tips from the cleaner. If<br />

the speed is too high, droplets may be dragged out of the cleaner and stick to<br />

the outside of the tips. Retract with is ignored if you are flushing DITI adapters.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 15 - 11

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!