15 - Script Commands<br />

Tip Alignment Check Command<br />

Before starting a script which contains the Tip Alignment Check command, you<br />

must place the Te-PS Sensor plate on one of the sites of the Te-PS Carrier. Check<br />

that the sensor plate is properly connected to the pipetting instrument using the<br />

supplied cable.<br />

In addition, you must put the Te-PS Sensor plate at the same grid and site position<br />

in <strong>Freedom</strong> <strong>EVOware</strong>’s Worktable Editor.<br />

If necessary, the Tip Alignment Check command can also be used to check the tip<br />

alignment with other carrier and labware types. The test should then be carried<br />

out with the Te-PS Sensor plate placed on the same carrier which will be used for<br />

pipetting. However, due to the mechanical design and other factors, the Te-PS<br />

Carrier can achieve a higher alignment accuracy than conventional carrier types.<br />

Fig. 15-20 Tip Alignment Check command<br />

The parameters of the Tip Alignment command are as follows:<br />

Alignment options panel<br />

Te-PS Sensor plate position<br />

Specify the location of the Te-PS Sensor plate by clicking on the sensor plate<br />

in the Worktable Editor.<br />

Script aborts if misalignment is more than ...<br />

Specify a maximum value in mm for tip misalignment above which the script<br />

should be aborted. The maximum allowed misalignment depends on the<br />

radius and shape of the wells in the labware that you are using. It should be<br />

less than well radius - tip radius. For 1536-well labware, an appropriate value<br />

for this parameter is 0.5 mm.<br />

If tip misalignment exceeds the specified value, the instrument must be recalibrated<br />

by authorized personnel using the Setup & Service software. See<br />

1.2 “Reference Documents”, 1-3 for information on this software.<br />