15 - Script Commands<br />

Worklist Command<br />

Tab. 15-1 Parameters for Aspirate and Dispense records in a worklist file (cont.)<br />

Parameter Range Explanation<br />

LiquidClass Max. 32 characters This optional parameter overwrites the liquid class<br />

specified in the Worklist command.<br />

TipType<br />

Reserved, must be omitted.<br />

TipMask 1 .. 128 This optional parameter specifies the tip you want to<br />

use. The tip number is bit-coded, i.e. Tip 1 = 1, Tip 2<br />

= 2, Tip 3 = 4; Tip 4 = 8 and so on. Please note that<br />

you can only use a tip which has been enabled with<br />

Tip Selection in the Worklist command. If you do not<br />

specify TipMask, <strong>Freedom</strong> <strong>EVOware</strong> uses one of the<br />

tips specified with Tip Selection in the Worklist command.<br />

ForcedRack-<br />

Type<br />

Max. 32 characters<br />

This optional parameter is the configuration name of<br />

the labware. If specified, it is used instead of the labware<br />

type which is found on the <strong>Freedom</strong> <strong>EVOware</strong><br />

worktable.<br />

MinDetected-<br />

Volume<br />

0 .. +7158278 Liquid volume in µl. If this optional parameter is<br />

specified and Liquid Level Detection is enabled in<br />

the selected liquid class, it is used by the LiHa to<br />

determine the minimum liquid height which must be<br />

available in the well. Otherwise a liquid detection<br />

error (not enough liquid) will be triggered.<br />

MinDetectedVolume specifies the usable volume of<br />

liquid. If the aspiration position in the liquid class<br />

specifies tip immersion (tip submerging), the actual<br />

liquid height in the well must be higher to allow<br />

immersion (i.e. the volume of liquid in the well must<br />

be more than MinDetectedVolume).<br />

15.24.2.2 Parameters for Reagent Distribution records<br />

The aspirate parameters for Reagent Distribution records are as follows:<br />

Tab. 15-2 Aspirate Parameters for the Reagent Distribution command<br />

Parameter Range Explanation<br />

SrcRackLabel<br />

Max. 32 characters<br />

User-defined label (name) which is assigned to the<br />

source labware.<br />

SrcRackID Max. 32 characters Source labware barcode.<br />

SrcRackType Max. 32 characters Source labware type (configuration name), e.g. “384<br />

Well, landscape”.<br />

SrcPosStart 1 .. number of wells First well to be used in the source labware.<br />

SrcPosEnd 1 .. number of wells Last well to be used in the source labware.<br />

The dispense parameters for Reagent Distribution records are as follows:<br />

15 - 42 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>