16 - Device Commands<br />

Commands for the Columbus Washer<br />

Aspiration Mode<br />

Choose the required aspiration mode. The available options are normal/<br />

normal, crosswise/crosswise, normal/crosswise and crosswise/normal.<br />

Crosswise aspiration should only be used for 96-well plates with flat bottomed<br />

wells. In this case, aspiration is carried out with the needles at two positions in<br />

each well (front edge and back edge). This minimizes the volume of residual<br />

liquids. For more information, please refer to the corresponding section in the<br />

operating manual of your washer.<br />

Wash mode<br />

Choose whether you want to use plate mode or strip mode. In plate mode,<br />

each process step is performed on all of the strips sequentially before<br />

proceeding to the next process step. The entire microplate is processed for<br />

the defined soaking time. In strip mode, the entire wash program is performed<br />

on a single strip (i.e. one column or row of the microplate, depending on wash<br />

orientation) before proceeding to the next strip. Each strip is processed for the<br />

defined soaking time before proceeding to the next strip. See also<br />

17.5 “Columbus Washer - Washing a Variable Number of Strips in the Plate”,<br />

17-15.<br />

Repetitions<br />

Specify the required number of repetitions of the washing procedure.<br />

Final Asp. Time<br />

Specify the required final aspiration time (range: 1 to 20 seconds). Final<br />

aspiration is used to remove remaining liquid from the wells after the washing<br />

procedure.<br />

Final Asp. Speed<br />

Specify the required final aspiration speed (range: 1 to 20 mm/s). This is the<br />

speed at which the needles are lowered into the wells as the liquid level drops.<br />

16.6.10.2 Wash Tab of the Columbus Washer - WashPlateEx Command<br />

The Wash tab is used to set additional washing parameters.<br />

Fig. 16-23 Columbus WashPlateEx command - Wash tab<br />