15 - Script Commands<br />

Serial Dilution Wizard<br />

How many steps should your dilution contain<br />

Specify the required number of steps (range: 2 to 24). Example: if you specify<br />

a dilution of 1:2 per step and a total of 3 steps, the destination plate(s) will all<br />

have three columns of wells with dilution factors of 1:2, 1:4 and 1:8. See also<br />

“Is the source concentration equal to the first step”.<br />

How many dilution series (plates)<br />

Specify how many destination plates you want to make (range: 1 to 16). All of<br />

the destination plates will be identical.<br />

Dilute the first transfer to the destination plate(s)<br />

Click No if you want the first column in the destination plate(s) to contain the<br />

undiluted sample. Example: if you choose No and specify a dilution of 1:2 per<br />

step and a total of 3 steps, the destination plate(s) will all have three columns<br />

of wells with dilution factors of “undiluted”, 1:2 and 1:4.<br />

Is the diluent already distributed<br />

Click Yes if the destination wells in the destination plate(s) already contain<br />

diluent. In this case, you must specify how much diluent is contained in each<br />

well, and the Serial Dilution wizard only needs to aspirate and dispense the<br />

appropriate quantities of the sample liquid. Tip: You can dispense diluent to<br />

the destination wells using the Replicates wizard.<br />

How much diluent is dispensed in each well: Specify how much diluent is<br />

already contained in each destination well.<br />

What final volume do you want to have<br />

Specify the required total volume (sample plus diluent) in each of the<br />

destination wells.<br />

Aspirate sample and diluent together<br />

Click Yes if you want to aspirate the sample directly after aspirating the<br />

diluent. Click No if you want to dispense the diluent into the destination wells<br />

before aspirating and dispensing the sample. Be aware of the risk of<br />

contamination if you choose Yes.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 15 - 71