9 - Specifying Your Instrument Setup<br />

Configuring Carriers<br />

9.5.5.1 Labware Carrier Attributes<br />

The attributes which you can configure for labware carriers are as follows:<br />

Tab. 9-25 Labware carrier attributes<br />

Field name Default value Description<br />

Grip orientation narrow (narrow side) This field specifies whether the RoMa should<br />

grip labware on this carrier on the narrow side<br />

or wide side. If you specify “narrow”, make<br />

sure that you have created a “Move Narrow”<br />

RoMa vector for this carrier (vice versa for<br />

“wide”) a) .<br />

Lid is allowed no (not allowed) Specify “yes” if a lid is allowed on labware<br />

which is on this carrier.<br />

Location to store lid<br />

Allowed robots<br />

(original position<br />

of labware)<br />

If lids are not allowed on labware which is on<br />

this carrier, the lid will be removed from the<br />

labware and moved to the location specified in<br />

this field. b)<br />

Comma separated list of all plate robots that<br />

can safely access this location (e.g. “RoMa,<br />

RoMa2”). The robot names are case sensitive.<br />

If this field is left empty, <strong>Freedom</strong><br />

<strong>EVOware</strong> assumes that all plate robots can<br />

safely access this location. The list may be<br />

necessary if e.g. the carrier is positioned at<br />

the extreme left or extreme right of the worktable.<br />

SafeMoveOffset 0 X-offset in mm for safe LiHa movement to and<br />

from the pipetting position (see 17.4 “Safe<br />

Pathways Feature”, 17-14).<br />

SafeMoveSpeed<br />

Speed to use for movements along safe paths<br />

in mm/s (see 17.4 “Safe Pathways Feature”,<br />

17-14).<br />

Safety 1 Only for the MCA384 DITI Carrier:<br />

1 = Close the locking mechanism when<br />

mounting DITIs from the MCA384 DITI Carrier.<br />

0 = Do not close the locking mechanism.<br />

ZTravel - Only if you choose Positioning with advanced<br />

Z-travel in the Configuration Tool (see<br />

8.4.2.11, 8-34). If the ZTravel parameter is<br />

non-zero, it is used for Z-travel height instead<br />

of the taught Z-travel height.<br />

a) If you specify “narrow” and the physical orientation of the carrier site is landscape (e.g. the “MP 3 Pos”<br />

carrier), the RoMa will grip the labware from the side of the worktable (with the grippers pointing in the<br />

direction X or -X), not from the front (vice versa for “wide”).<br />

b) Broadly speaking, the lid attributes in the labware definition specify how to grip the lid; the lid attributes in<br />

the carrier definition specify where to store the lid if it is removed.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 9 - 49