18 - Error Handling<br />

List of Error Messages<br />

Tab. 18-1 List of error messages (cont.)<br />

Error ID Message Text Reason Corrective Measure<br />

EVO_EVO_013_021 Invalid operand PosID firmware command<br />

error<br />

EVO_EVO_013_022 Invalid number of operands PosID firmware command<br />

error<br />

EVO_EVO_013_023 Invalid device address PosID firmware command<br />

error<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

EVO_EVO_013_024 Door already open n.a. n.a.<br />

EVO_EVO_013_025 Hall sensors not calibrated n.a. Field Service Engineer:<br />

Calibrate hall sensors<br />

EVO_EVO_013_026 Pump not defined MPO/SPO firmware command<br />

error<br />

Contact your local service<br />

organization<br />

EVO_EVO_013_027 LICOS not calibrated n.a. Field Service Engineer:<br />

Calibrate LICOS<br />

EVO_EVO_013_028 No access to EEPROM MPO/SPO firmware command<br />

error<br />

Contact your local service<br />

organization<br />

EVO_EVO_013_029 Pressure sensor not calibrated n.a. Field Service Engineer:<br />

Calibrate pressure sensor<br />

EVO_EVO_013_030 Command overflow of sub device MPO/SPO firmware command<br />

error<br />

EVO_EVO_013_031 Pump stalled MPO/SPO firmware command<br />

error<br />

EVO_EVO_013_037 Initialization error Diluter firmware command<br />

error<br />

EVO_EVO_013_038 Invalid command Diluter firmware command<br />

error<br />

EVO_EVO_013_039 Invalid operand Diluter firmware command<br />

error See<br />

18.1.7, 18-7<br />

EVO_EVO_013_040 Invalid command sequence Diluter firmware command<br />

error<br />

EVO_EVO_013_041 Dilutor not connected. Diluter firmware command<br />

error<br />

EVO_EVO_013_042 Error in EEPROM Diluter firmware command<br />

error<br />

EVO_EVO_013_043 Device not initialized Diluter firmware command<br />

error<br />

EVO_EVO_013_044 Plunger overload Defective valve<br />

Clogged liquid system<br />

See 18.1.7, 18-7<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

Check diluter<br />

Check liquid system<br />

EVO_EVO_013_045 Valve blocked Defective valve Replace valve<br />

18 - 46 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>