15 - Script Commands<br />

Te-Fill Switch Valve Command<br />

Channel / Liquid<br />

If the Te-Fill is fitted with the 6-way valve option, choose the channel (liquid<br />

container) you want to use. The pull-down list is not available if the 6-way<br />

valve is not present and/or not activated.<br />

During runtime, the pipetting instrument will select the correct liquid channel<br />

automatically - you do not need to use the Te-Fill Switch Valve command for<br />

this purpose.<br />

Speed<br />

Specify the required priming speed for each tip. The available range depends<br />

on how many tips you have selected. The Te-Fill pump has a range of 8 to 80<br />

ml/min. If you have selected e.g. four tips, the range for this field is 2 to 20 ml/<br />

min.<br />

15.17 Te-Fill Switch Valve Command<br />

The Te-Fill option uses a bi-directional pump instead of the diluter to aspirate and<br />

dispense large volumes of liquid, for example buffer solutions.<br />

The Te-Fill can be optionally equipped with a 6-way valve, which allows you to<br />

switch between six different liquid containers. If fitted, the 6-way valve must be<br />

activated in the Configuration Tool (see 8.4.2.4 “Te-Fill option”, 8-30). Also use<br />

the Configuration Tool to assign the names of the liquids you want to use to the<br />

channels of the 6-way valve (see 8.4.2.6 “6-way valve for Te-Fill”, 8-31). The<br />

assigned names will then be shown in this command in the Channel / Liquid pulldown<br />

list.<br />

With the Te-Fill Aspirate, Dispense and Prime commands, the pipetting instrument<br />

will switch during runtime from using the diluter to using the bi-directional pump<br />

and switch back again after pipetting - you do not need to use the Te-Fill Switch<br />

Valve command for this purpose.<br />

The Te-Fill Switch Valve command is used for special applications which require<br />

the Te-Fill liquid path to be selected but without aspirating or dispensing.<br />

Fig. 15-17 Te-Fill Switch Valve command<br />

The parameters of the Te-Fill Switch Valve command are as follows:<br />