10 - Configuring Liquid Classes<br />

MCA96 Liquid Classes<br />

10.3.2 MCA96 Dispense Parameters<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses these parameters for the MCA96 pipetting head when<br />

dispensing or mixing.<br />

Fig. 10-8 MCA96 Dispense parameters<br />

The parameters of this dialog box are as follows (some of the parameters must be<br />

specified separately for single pipetting mode and multi pipetting mode):<br />

Dispense Speed<br />

Specify the dispense speed in µl/s. A high speed (400-600 µl/s) is needed to<br />

achieve a proper free dispense. Contact dispense requires a lower speed (in<br />

the same range as the aspirate speed).<br />

Breakoff Speed<br />

You cannot specify a breakoff speed for the MCA96.<br />

Delay<br />

Specify the delay after dispensing.<br />

Trailing Airgap after each Dispense<br />

You can automatically aspirate a trailing airgap after each dispense operation<br />

in order to avoid droplets forming on the tip. The trailing airgap parameters are<br />

specified in the Aspirate dialog box.<br />

Use Pinch Valve<br />

The MCA96 does not have a pinch valve.<br />

Use Liquid Detection<br />

The MCA96 does not support liquid level detection (or emulated liquid level<br />

detection) for dispensing.<br />