10 - Configuring Liquid Classes<br />

LiHa Liquid Classes for Fixed Tips and DiTis<br />

10.1.2 Type Parameters for Liquid Classes<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses these parameters to select the best suited sub-class for<br />

a specified tip type and pipetting volume.<br />

Fig. 10-3 Type parameters for liquid classes<br />

Tip Type<br />

Select the tip type. The following tip types are available:<br />

– <strong>Standard</strong> tips<br />

– DITI (disposable tips)<br />

– Std. & Low Vol. (standard tips with the low volume option)<br />

– DITI & Low Vol. (disposable tips with the low volume option)<br />

– Te-MO Fixed tips (96 tips)<br />

– Te-MO DITI tips (96 tips)<br />

– Te-MO 384 (384-tip pipetting head)<br />

– MCA96 Fixed (Multi-Channel Arm with fixed tips)<br />

– MCA96 DITI (Multi-Channel Arm with disposable tips)<br />

– MCA384 Fixed (Multi-Channel Arm with fixed tips)<br />

– MCA384 DITI (Multi-Channel Arm with disposable tips)<br />

The following options are only intended for research applications: Low<br />

Volume, Te-MO tips (all types).<br />

Volume Range<br />

If you check the Use for All Volumes checkbox, the aspirate, dispense and<br />

calibration parameters for this sub-class are valid for all pipetting volumes.<br />

If this box is not checked, you must specify the minimum volume and the<br />

maximum volume. <strong>Freedom</strong> <strong>EVOware</strong> will then use this sub-class during<br />

pipetting if the following expression is true:<br />

Minimum Volume ≤ pipetting volume < Maximum Volume<br />

10.2 LiHa Liquid Classes for Fixed Tips and DiTis<br />

The LiHa (Liquid Handling Arm) is a robot arm on the pipetting instrument which is<br />

mounted on the X-slide and typically holds four or eight pipetting tips.<br />

10.2.1 LiHa Aspirate Parameters (Fixed Tips and DiTis)<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses these parameters when aspirating or mixing with LiHa<br />

fixed tips or DITIs. The illustrations at the top right of the dialog box show<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 10 - 7