18 - Error Handling<br />

List of Error Messages<br />

Tab. 18-1 List of error messages (cont.)<br />

Error ID Message Text Reason Corrective Measure<br />

EVO_EVO_014_006<br />

There are schedules stored for this<br />

process. If the process is changed<br />

these schedules will be lost.<br />

To keep the original version with it's<br />

schedules please use 'Save As' to<br />

store the changed process under a<br />

new name.<br />

Save process under same name and<br />

loose schedules?<br />

n.a.<br />

n.a.<br />

EVO_EVO_015_000 Error opening Named Pipe Named Pipe could not be<br />

opend in “Command”<br />

command<br />

• Check pipe name provided<br />

in command<br />

• Wait until pipe is available<br />

EVO_EVO_016_002 Enter Barcode (PosID-2) Refer to 18.1.16, 18-16<br />

EVO_EVO_016_003 Barcode Error (PosID-2) Refer to 18.1.12, 18-13<br />

EVO_EVO_016_004 Clot Error Refer to 18.1.2, 18-2<br />

EVO_EVO_016_005 Dilutor Overflow Refer to 18.1.6, 18-6<br />

EVO_EVO_016_006 Dilutor Error Refer to 18.1.7, 18-7<br />

EVO_EVO_016_007 DITI Already Mounted Refer to 18.1.11, 18-12<br />

EVO_EVO_016_008 Error Fetching DITIs Refer to 18.1.8, 18-9<br />

EVO_EVO_016_009 Te-MO DITI Error Refer to 18.1.22, 18-22<br />

EVO_EVO_016_010 Lost DITI Refer to 18.1.10, 18-11<br />

EVO_EVO_016_011 Doorlock Error Refer to 18.1.19, 18-19<br />

EVO_EVO_016_023 Liquid Arrival Check Error Refer to 18.1.5, 18-5<br />

EVO_EVO_016_024 Liquid Detection Error Refer to 18.1.1, 18-1<br />

EVO_EVO_016_036 Grip Error Refer to 18.1.20, 18-20<br />

EVO_EVO_016_040 Instrument Error dialog box Refer to 18.1.18, 18-18<br />

EVO_EVO_016_042 Te-Stack Error Refer to 18.1.21.1, 18-21<br />

EVO_EVO_016_043 Vacuum Waste Container Refer to 18.1.26, 18-27<br />

EVO_EVO_016_048 WinWash Liquid Error Refer to 18.1.25, 18-26<br />

EVO_EVO_016_049 WRC/Te-MO Error Refer to 18.1.23, 18-23<br />

EVO_EVO_016_052 Clot Error (PMP) Refer to 18.1.3, 18-3<br />

EVO_EVO_016_053 PMP Instrument Error Refer to 18.1.4, 18-4<br />

EVO_EVO_017_000<br />

A Te-MO has been detected but is<br />

not on the worktable.<br />

Evoware now prohibits LiHa access<br />

to the rear of the instrument.<br />

n.a.<br />

n.a.<br />