10 - Configuring Liquid Classes<br />

Te-MO 96 Liquid Classes<br />

Dispense Position<br />

Specify the Z-position and the Z-offset of the tips when dispensing. A positive<br />

value for the offset lowers the tips.<br />

If you choose a dispense position with tracking, the tips move upwards as the<br />

liquid level rises in the wells during dispensing. This option can be used e.g. to<br />

ensure minimal tip contamination by limiting the tip immersion depth.<br />

You can also choose the X-position when dispensing as follows:<br />

– Center: as specified for the labware which is used<br />

– Left: left of center; the offset is specified with X or Y offset while pipetting<br />

for the labware which is used<br />

– Right: right of center; the offset is specified with X or Y offset while<br />

pipetting for the labware which is used<br />

– Left / Right: left of center for odd columns, right of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

– Right / Left: right of center for odd columns, left of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

You can also choose the Y-position when dispensing as follows:<br />

– Center: as specified for the labware which is used<br />

– Rear: rear of center; the offset is specified with X or Y offset while pipetting<br />

for the labware which is used<br />

– Front: front of center; the offset is specified with X or Y offset while<br />

pipetting for the labware which is used<br />

– Rear / Front: rear of center for odd columns, front of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

– Front / Rear: front of center for odd columns, rear of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

By combining an X offset and a Y offset, there are nine possible pipetting<br />

positions (including the center position with no offsets).<br />

Tip Touching<br />

Activate the tip touching function if required and specify the direction of the<br />

movement. The amount of movement is specified with Tip-touching distance<br />

for the labware which is used.<br />

Tip touching is a special movement when dispensing a liquid. When tip<br />

touching is activated, <strong>Freedom</strong> <strong>EVOware</strong> dispenses to the destination<br />

container in the normal way, but then moves the tips to the side of the well and<br />

back before retracting them. This helps to remove any droplets which may be<br />

adhering to the tips.<br />

Speed: Specify the tip touching speed.<br />

Delay after touching: Specify the delay between the tip touching the well and<br />

moving back to the middle of the well.<br />

Mix after Dispense<br />

Check this box if you want to mix the liquid after dispensing it. You can specify<br />

the number of mix cycles (range: 1 to 99), the mix volume (max. 2500 µl), the<br />

Z-position and the Z-offset. A positive value for the offset lowers the tips.<br />

Retract Tips to<br />

Specify the Z-position and the Z-offset to which the tips should retract (move<br />

upwards) after dispensing. A positive value for the offset lowers the tips.<br />

10 - 36 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>