18 - Error Handling<br />

List of Error Messages<br />

Tab. 18-1 List of error messages (cont.)<br />

Error ID Message Text Reason Corrective Measure<br />

EVO_EVO_001_001<br />

Error in configuration file in line<br />

<br />

One of the configuration<br />

files is not accessible for<br />

reading<br />

• Reinstall <strong>EVOware</strong><br />

• Restore backup<br />

EVO_EVO_001_002 Error writing configuration file One of the configuration<br />

files is not accessible for<br />

writing<br />

EVO_EVO_001_003 Incompatible CFG File A configuration file is currupt<br />

• Reinstall <strong>EVOware</strong><br />

• Restore backup<br />

• Reinstall <strong>EVOware</strong><br />

EVO_EVO_001_004<br />

Old version of configuration file!<br />

Please define all elements new.<br />

A configuration file stems<br />

from an older version of<br />

<strong>EVOware</strong><br />

• Reinstall <strong>EVOware</strong><br />

• Use current version of<br />

file<br />

EVO_EVO_001_005 LIHA not defined A configuration file is currupt<br />

• Configure LiHa in Configuration<br />

• Reinstall <strong>EVOware</strong><br />

EVO_EVO_002_000<br />

Diluter broken!<br />

Switching Diluter off!<br />

Diluter error/defective<br />

Check/replace diluter<br />

EVO_EVO_002_001<br />

DITI not dropped for tip.<br />

Retry?<br />

Problem with DiTi<br />

DiTi drop mechanism<br />

defective<br />

User new DiTi<br />

Check DiTi drop mechanism<br />

EVO_EVO_002_002<br />

Internal error.<br />

Different pressure sensitivities used<br />

in one liquid detection action<br />

<strong>Software</strong> error<br />

Contact your local service<br />

organization<br />

EVO_EVO_002_003<br />

Internal error.<br />

Different liquid characteristics used<br />

in one liquid detection action<br />

<strong>Software</strong> error<br />

Contact your local service<br />

organization<br />

EVO_EVO_002_004<br />

No MPO found<br />

LICOS not installed<br />

No MPO installed<br />

MPO defective<br />

Check/replace MPO<br />

Install LICOS<br />

EVO_EVO_002_005<br />

No way to LiHa target<br />

found!\\Evoware switches into offline<br />

mode.<br />

Inconsistent process definition<br />

Check process definition<br />

EVO_EVO_002_006 Internal Error: Illegal Move <strong>Software</strong> error Contact your local service<br />

organization<br />

EVO_EVO_003_001<br />

Error unloading device drivers: <br />

Inconsistent installation<br />

Reinstall <strong>EVOware</strong><br />

EVO_EVO_003_002<br />

Error shutting down devices:<br />

<br />

Inconsistent installation<br />

Reinstall <strong>EVOware</strong><br />

EVO_EVO_004_000<br />

Maximum tip-misalignment<br />

exceeded !<br />

Script execution will be aborted.<br />

Tips misaligned/bent<br />

Align/replace tips<br />

EVO_EVO_005_000 Error starting processes: Process not executable Modify process<br />

EVO_EVO_005_001<br />

Your process completed successfully<br />

!<br />

n.a.<br />

n.a.<br />