8 - Configuration Tool<br />

Device Driver Configuration<br />

Number of slots per tower<br />

Choose the number of slots which are available per physical incubator (4 or<br />

6). If the logical incubator extends over more than one physical incubator, all<br />

of them must have the same number of slots.<br />

Temperature<br />

Specify the temperature of the logical incubator (range, model MIO1: 25° to<br />

47°C; range, model MIO2: 25° to 65°C). The same temperature is used for all<br />

of the slots of the logical incubator. Specify 0 to disable the heater. To ensure<br />

that the incubator reaches and maintains a stable temperature, make sure<br />

that the incubator doors are not left open.<br />

Note: For special applications, you can use the SetTemperature command to<br />

assign a different temperature to individual slots during script or process runtime.<br />

Shaker speed<br />

Specify the speed at which the labware should be shaken during incubation<br />

(range, model MIO1: 1 to 10.5 Hz; range, model MIO2: 1 to 8.5 Hz). Specify 0<br />

to disable shaking. Shaking requires the incubator shaker option.<br />

Towers panel<br />

Number of Towers<br />

Choose how many physical incubators are used by this logical incubator<br />

(maximum 8). Depending on your choice, additional tabs will appear (1, 2, 3<br />

etc.).<br />

Device address<br />

For each of the Towers tabs (physical incubators), choose the device address<br />

which is associated with this logical incubator. Choose the address as<br />

“hardware address plus 1”. On heated incubators, the hardware address is set<br />

with the blue address switch at the back. Each physical device (regardless of<br />

device type) must have a unique hardware address.<br />

Examples<br />

Example 1: Two logical incubators (Incubator1, Incubator2) each have three slots<br />

and share the same physical incubator (tower 1). Incubator1 starts at slot 1 (the<br />

topmost slot) and has a temperature of 35 °C; Incubator2 starts at slot 4 and has a<br />

temperature of 40 °C.<br />

Example 2: One logical incubator has 7 slots and starts at slot 4 of tower 1. The<br />

last used slot is then slot 3 of tower 2.<br />

Additional settings in the configuration file for the incubator driver<br />

Two of the parameters in the [Settings] section of the configuration file for the<br />

incubator driver allow you to adjust the warning level if the incubator temperature<br />

differs from the configured value.<br />

• The configuration file is located in the <strong>Freedom</strong> <strong>EVOware</strong> installation directory<br />

and has the name Te-Incubator1.ini (for the first heated incubator) to Te-<br />

Incubator3.ini (for the third heated incubator).<br />

• The parameter TempRange specifies the threshold for the warning. The<br />

incubator driver interrogates the actual incubator temperature once per<br />

minute and a warning is output if the temperature differs from the configured<br />

temperature by more than the specified range (either hotter or colder).<br />

TempRange is specified as an unsigned integer in °C.<br />

• The parameter TempWarnDelay specifies a time delay after inserting the<br />

labware in the incubator and before the first warning is output. This allows<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 8 - 57