15 - Script Commands<br />

MCA96 Commands<br />

Volume<br />

Specify the volume that you want to aspirate for the mixing process. The same<br />

volume is pipetted by all of the tips. The maximum allowed volume is the<br />

plunger or DITI capacity (<strong>Freedom</strong> <strong>EVOware</strong> does not automatically handle<br />

volumes that exceed the plunger or DITI capacity).<br />

Cycles<br />

Specify how often you want to repeat the mixing process.<br />

Compartment in well<br />

Some special labware types (e.g. for protein crystallography) have wells with<br />

several compartments (cavities). <strong>Freedom</strong> <strong>EVOware</strong> can pipette a different<br />

liquid into each of the compartments. Each well typically has one main<br />

compartment and one or more satellite compartments which are directly<br />

adjacent and are often smaller in size. Choose the compartment which you<br />

want to use for pipetting (the main compartment is normally compartment 1).<br />

Well compartments are specified in the labware configuration. This field is<br />

inactive (gray) if the well only has one compartment.<br />

Location<br />

Specify the required labware by clicking on the labware in the Worktable<br />

Editor. You must first put an appropriate labware item in the Worktable Editor<br />

at the required position. The gray (protected) field above the location panel<br />

then shows the label (name) of the labware that you have selected (it shows a<br />

generic name if you have not assigned a label).<br />

If the MCA96 pipetting head has less tips than the number of wells in the<br />

labware, you must click in the Location panel to select the set of wells you<br />

want to use. This is not necessary if the pipetting head has the same number<br />

of tips as the number of wells in the labware. Your selection will be shown in<br />

green.<br />

15.68.4 MCA96 Wash Command<br />

The MCA96 Wash command is used to wash MCA96 tips or MCA96 DITIs using<br />

the specified MCA96 wash block. The MCA96 wash block is intended for 96-tip<br />

MCA96 tip blocks and 96-tip DITi pipetting heads and is connected to the MCA96<br />

Wash & Refill Center (WRC) through tubing. The MCA96 hardware supports up to<br />

two MCA96 wash blocks (each one with its own control unit). You can only put<br />

MCA96 wash blocks on the carrier type “MCA96 Carrier”.<br />

15 - 164 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>