10 - Configuring Liquid Classes<br />

LiHa Liquid Classes for Fixed Tips and DiTis<br />

Trailing Airgap after each Dispense<br />

You can automatically aspirate a trailing airgap after each dispense operation<br />

to avoid droplets forming on the tips. The trailing airgap parameters for<br />

dispensing are specified in the Aspirate dialog box.<br />

Use Pinch Valve<br />

This is used to activate the Low Volume option if it is fitted to your pipetting<br />

instrument.<br />

Use Liquid Detection<br />

Check this box if you want to use capacitative liquid level detection (cLLD)<br />

when dispensing. cLLD detects the liquid surface through a change of the<br />

electrical capacitance of the pipetting tip when the tip enters the liquid. Liquid<br />

level detection when dispensing can be used e.g. to ensure that the tips<br />

remain immersed in the liquid during dispensing.<br />

cLLD is carried out with the detection mode, liquid conductivity, detection<br />

speed and double distance settings which you specified on the Global dialog<br />

box for the liquid class.<br />

cLLD must not be used if you are carrying out septum piercing (the septum<br />

interferes with the capacitive detection system).<br />

Dispense Position<br />

Specify the Z-position and the Z-offset of the tips when dispensing. A positive<br />

value for the offset lowers the tips. You can only choose a Z-position relative<br />

to the liquid level if you have activated liquid level detection.<br />

If you choose a dispense position with tracking, the tips will move upwards as<br />

the liquid level rises in the wells during dispensing. This option can be used<br />

e.g. to ensure minimal tip contamination by limiting the tip immersion depth.<br />

You can also choose the X-position when dispensing as follows:<br />

– Center: as specified for the labware which is used<br />

– Left: left of center; the offset is specified with X or Y offset while pipetting<br />

for the labware which is used<br />

– Right: right of center; the offset is specified with X or Y offset while<br />

pipetting for the labware which is used<br />

– Left / Right: left of center for odd columns, right of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

– Right / Left: right of center for odd columns, left of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

You can also choose the Y-position when dispensing as follows:<br />

– Center: as specified for the labware which is used<br />

– Rear: rear of center; the offset is specified with X or Y offset while pipetting<br />

for the labware which is used<br />

– Front: front of center; the offset is specified with X or Y offset while<br />

pipetting for the labware which is used<br />

– Rear / Front: rear of center for odd columns, front of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

– Front / Rear: front of center for odd columns, rear of center for even<br />

columns; the offset is specified with X or Y offset while pipetting for the<br />

labware which is used<br />

By combining an X offset and a Y offset, there are nine possible pipetting<br />

positions (including the center position with no offsets).<br />