15 - Script Commands<br />

<strong>Manual</strong> Tip Alignment Command<br />

with the Te-PS Sensor plate placed on the same carrier which will be used for<br />

pipetting. However, due to the mechanical design and other factors, the Te-PS<br />

Carrier can achieve a higher alignment accuracy than conventional carrier types.<br />

The following dialog box is shown when you insert the <strong>Manual</strong> Tip Alignment<br />

command in your script:<br />

Fig. 15-28 Sensorplate Position dialog box<br />

Click on the Te-PS Sensor plate in the Worktable Editor or specify the grid and site<br />

position manually and click OK to start the tip alignment procedure.<br />

A dialog box (not shown) warns you that the LiHa will now move to the sensor<br />

plate position. If your instrument is fitted with doorlocks, for safety reasons the<br />

doorlocks will be locked before moving the LiHa. Click OK to continue to the Tip<br />

Alignment Results dialog box or click Cancel to cancel tip alignment.<br />

15.27.1 Tip Alignment Results<br />

This dialog box is used by the <strong>Manual</strong> Tip Alignment function to display the<br />

misalignment (if any) of the Liquid Handling Arm (LiHa):<br />

Fig. 15-29 Tip Alignment Results dialog box<br />

Click OK to start the test. After the test is finished, the dialog box shows the<br />

misalignment in mm for each of the tips.<br />

If the misalignment exceeds the maximum allowed value for any of the tips, the<br />

instrument must be re-calibrated by authorized personnel using the Setup &<br />

Service software (see 1.2 “Reference Documents”, 1-3 for information on this<br />

software).<br />

The maximum allowed misalignment depends on the radius and shape of the<br />

wells in the racks that you are using. It should be less than well radius - tip radius.<br />