8 - Configuration Tool<br />

Device Driver Configuration<br />

Miscellaneous panel<br />

Initialization timeout<br />

Specify the maximum duration of the device initialization procedure. An error<br />

message will be output if initialization is not completed within this time.<br />

Advanced DITI handling<br />

This checkbox is not relevant for the MCA384 and should be ignored.<br />

Allow stacker commands in script<br />

This setting is only relevant for <strong>Freedom</strong> <strong>EVOware</strong> Plus. When you create a<br />

pipetting script for the MCA384 in an MCA384 Pipette process step, the<br />

Commands section of the Control Bar only shows relevant and/or permissible<br />

script commands. Te-Stack commands are not shown because stacker<br />

actions are normally part of a robot move or programmed using a separate Te-<br />

Stack process step. Check this checkbox to show Te-Stack commands in the<br />

Control Bar when you create a pipetting script for the MCA384. This is<br />

intended for special applications, above all for getting nested DITIs for the<br />

MCA384 from the stacker. The Te-Stack commands are only shown if you<br />

have installed the Te-Stack driver.<br />

In case of script error panel<br />

Request system pause<br />

Check this checkbox if you want the driver to request a system pause if a<br />

script error is detected during command execution.<br />

Move panel<br />

Grip Detection<br />

Specify whether grip detection should be used.<br />

Home after move<br />

Specify whether the MCA384 should be moved to its home position after each<br />

move. This may be necessary for some worktable layouts, but will reduce<br />

system performance.<br />

Regrip location<br />

This setting is analogous to the regrip location for the RoMa. See<br />

8.5.23.2 “Plate Robot Configuration - Settings Tab”, 8-86, Regrip location.<br />