15 - Script Commands<br />

Worklist Command<br />

The Set DITI Type record can only be used at the very beginning of the worklist or<br />

directly after a Break record. A Break record always resets the DITI type to the<br />

type selected in the Worklist command. Accordingy, if your worklist contains a<br />

Break record, you may need to specify the Set DITI Type record again.<br />

The Comment record (ignored by <strong>Freedom</strong> <strong>EVOware</strong>) has the following<br />

structure:<br />

C;Comment<br />

The Reagent Distribution record has the following structure:<br />

R;AspirateParameters;DispenseParameters;Volume;LiquidClass;NoOfDitiRe<br />

uses;NoOfMultiDisp;Direction[;ExcludeDestWell]*<br />

where:<br />

AspirateParameters =<br />

SrcRackLabel;SrcRackID;SrcRackType;SrcPosStart;SrcPosEnd;<br />

and<br />

DispenseParameters =<br />

DestRackLabel;DestRackID;DestRackType;DestPosStart;DestPosEnd;<br />

See 15.24.2.2 “Parameters for Reagent Distribution records”, 15-42 for details<br />

of the parameters.<br />

The Reagent Distribution record specifies the aspiration and dispensing<br />

parameters for all of the tips you have chosen (and dispenses the same volume in<br />

all of the chosen wells in the destination labware).<br />

The Reagent Distribution record has the following advantages when compared to<br />

separate Aspirate and Dispense records:<br />

• Pipetting is automatically optimized during script runtime for the number of tips<br />

which are actually fitted to the LiHa.<br />

• Pipetting is automatically adapted to special LiHa configurations such as the 8<br />

Plus 1 Access option of the <strong>Freedom</strong> EVO 75 pipetting instrument - it is not<br />

necessary to create a custom, LiHa-specific worklist.<br />

• Worklists are shorter and easier to set up manually using an editor - less risk<br />

of programming errors.<br />

For some examples, see 15.24.6 “Worklist Examples for the Reagent Distribution<br />

Record”, 15-45.<br />

15.24.2.1 Parameters for Aspirate and Dispense records<br />

The parameters for Aspirate and Dispense records are as follows:<br />

Tab. 15-1 Parameters for Aspirate and Dispense records in a worklist file<br />

Parameter Range Explanation<br />

RackLabel Max. 32 characters User-defined label (name) which is assigned to the<br />

labware.<br />

RackID Max. 32 characters Labware barcode.<br />

RackType Max. 32 characters Labware type (configuration name), e.g. “384 Well,<br />

landscape”.<br />

Position 1 .. number of wells Well position in the labware. The position starts with<br />

1 and increases from rear to front and left to right.<br />

TubeID Max. 32 characters Tube barcode.<br />

Volume 0 .. +7158278 Pipetting volume in µl.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 15 - 41