15 - Script Commands<br />

MCA384 Commands<br />

You can pick up one row or one column of DITIs if the DITI rack is on the<br />

MCA384 System Carrier but you cannot pick up all 384 DITIs. The latter<br />

option requires a locking mechanism for mechanical reasons which is not<br />

provided on the MCA384 System Carrier.<br />

If you pick up one row or one column of DITIs, the Pick up parameter in the<br />

Get DITIs command lets you choose whether to mount them on the first or last<br />

column (left or right side) or the first or last row (rear or front side) of the<br />

pipetting head. The Position parameter then lets you specify the row or<br />

column in the DITI rack from where to get the DITIs. This requires a relative<br />

movement of the pipetting head.<br />

Please note the following mechanical limitations: If you want to pick up e.g.<br />

DITIs on the first column (left side) of the pipetting head, you must pick them<br />

up starting with the last column (right side) of the DITI rack. If you pick up<br />

DITIs from column 23 and column 24 still contains DITIs, you will pick up two<br />

columns of DITIs instead of one column. Analogous limitations apply to<br />

picking up DITIs on the right side of the pipetting head and picking up single<br />

rows of DITIs on the front and rear of the pipetting head.<br />

Adapter DiTi MCA384<br />

You can only pick up DITIs if the DITI rack is on the MCA384 DITI Carrier. You<br />

can only pick up all 384 DITIs. The MCA384 DITI Carrier does not allow<br />

relative movements of the pipetting head.<br />

Adapter 96 DiTi MCA384<br />

You can pick up all 96 DITIs out of a box of 384 DITIs on any of the carrier<br />

types that allow the MCA384 DITI rack.<br />

You can pick up one row or one column of DITIs out of a box of 384 DITIs if<br />

the DITI rack is on the MCA384 System Carrier. This requires a relative<br />

movement of the pipetting head which is not possible for mechanical reasons<br />

on the MCA384 DITI Carrier.<br />

Adapter 96 DiTi 4to1 MCA384<br />

The limitations for this head adapter are the same as “Adapter 96 DiTi<br />

MCA384”.<br />

Adapter QC MCA384<br />

This special head adapter is used to protect the pipetting head during<br />

transport and storage and is not intended for pipetting.<br />

15.69.9 MCA384 Drop DITIs Command<br />

The MCA384 Drop DITIs command is used to return the disposable tips of the<br />

multi channel arm to a specified position on the worktable of the pipetting<br />

instrument.<br />

The MCA384 Drop DITIs command also allows you to drop one row or one<br />

column of DITIs in to a box of 384 DITIs. See 15.69.9.1 “Limitations for different<br />

head adapter types (dropping DITIs)”, 15-197 for more information.<br />