16 - Device Commands<br />

Commands for the Magnetic Bead Separator<br />

Fig. 16-72 Magnetic Bead Separation Execution command<br />

To add a command to the separation script, click one of the Commands buttons<br />

after specifying the command parameters to the right of the button. To change the<br />

order of execution of the commands, select a command from the list and click the<br />

Up or Down buttons. To delete a command from the list, select the command and<br />

click Delete.<br />

The following commands are available:<br />

Magnetic Mix<br />

This command mixes the magnetic beads with the sample liquid by applying<br />

the magnet block to alternate sides of the carrier the specified number of<br />

times (Cycles) with the specified waiting time at each of the two positions. In<br />

addition to moving from left to right and back, the magnet block also moves up<br />

and down to follow an approximately rectangular path.<br />

Z-Position<br />

Specify the required Z-position (highest position) of the magnet block for the<br />

Magnetic Mix command (range: 1 to 31 movement steps).<br />

z = 31 is the highest possible position and results in the maximum up and<br />

down movement of the magnet block during the Magnetic Mix command. z = 1<br />

is the lowest possible position and corresponds to the Dispense position. In<br />

this case, the Magnetic Mix command will only move the magnet block from<br />

left to right and back and not up and down.<br />

Specify a smaller value for Z-position if you only have a small volume of liquid<br />

in the labware. This ensures that the upwards movement of the magnet block<br />

will not drag the magnetic beads above the liquid surface. This makes it easier<br />

to aspirate the supernatant or wash liquid with the LiHa in a following step.<br />

This field is only shown if you have checked Support Z-positioning in the<br />

configuration of the Te-MagS driver. This feature requires special Te-MagS<br />

hardware.<br />

Incubation<br />

This command incubates the labware on the heater for the specified time.<br />