18 - Error Handling<br />

Error Dialogs<br />

Send<br />

This button lets you send a firmware command directly to the instrument and<br />

is only intended for specialists. Specify the required command in the<br />

Command field and click Send. For example, “A1RPZ0” returns the current Z<br />

position of each of the LiHa tips. See the respective instrument manual for a<br />

list of firmware commands. Firmware commands are nearly always written<br />

uppercase.<br />

Error: All firmware commands terminate with an error code. After clicking the<br />

Send button, the error code is shown in this (protected) field. The error code is<br />

0 if the command was executed with no errors.<br />

Answer: Some firmware commands terminate with one or more return<br />

parameters (numeric values) separated by commas. After clicking the Send<br />

button, the return parameters are shown in this (protected) field.<br />

WARNING<br />

Risk of injury to operating personnel or damage to the equipment<br />

• Take care when sending firmware commands. Sending the wrong commands<br />

can damage the instrument or objects in its workspace, hurt yourself or other<br />

people.<br />

18.1.24 WRC Errors<br />

The Wash & Refill Center is a supporting device for the Te-MO which is connected<br />

to the Te-MO wash block through tubing. The following two errors can be output<br />

by the Wash and Refill Center. See 18.1.23 “WRC/Te-MO Error”, 18-23 for<br />

information on additional Te-MO WRC errors.<br />

18.1.24.1 WRC Empty Error<br />

The following dialog box is shown during a Te-MO wash cycle if the wash block of<br />

the Wash & Refill Center (WRC) is empty. You will be prompted to refill the wash<br />

block:<br />

Fig. 18-25 Wash & Refill Center error dialog box<br />

The buttons in the dialog box are as follows:<br />

OK<br />

The pumps will be started and will continue to run until the liquid reaches the<br />

specified upper level or the user stops the pumps.<br />

18 - 24 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>