8 - Configuration Tool<br />

Device Settings<br />

8.4.2.6 6-way valve for Te-Fill<br />

The Te-Fill option uses a bi-directional pump instead of the diluter to aspirate and<br />

dispense large volumes of liquid, for example buffer solutions. The Te-Fill option<br />

can be optionally provided with a 6-way valve, which allows you to switch between<br />

six different liquid containers.<br />

Specify the names of the liquids you want to use to the channels of the 6-way<br />

valve. Click to add a new entry to the list. Click to delete the currently<br />

selected entry. Click or to change the sort order of the entries.<br />

The assigned names will then be shown in the Te-Fill pipetting commands in the<br />

Channel / Liquid pull-down list.<br />

See also see 8.4.2.4 “Te-Fill option”, 8-30.<br />

8.4.2.7 Doorlocks<br />

<strong>Freedom</strong> <strong>EVOware</strong> supports up to four electrically-operated doorlocks which are<br />

fitted to the pipetting instrument and the doorlocks are arranged in groups. This<br />

function is used to assigned each of the doorlocks to one of the four available<br />

groups.<br />

The doorlocks of all groups are locked automatically when starting or continuing a<br />

script. The doorlocks of all groups are unlocked automatically when finishing,<br />

pausing or aborting a script. See 2.6 “Safety Features of the Pipetting Instrument”,<br />

2-4.<br />

Doorlock panel<br />

The <strong>Freedom</strong> EVO pipetting instrument is always fitted with two doorlocks<br />

(Doorlock 1 and Doorlock 2) and they are permanently assigned to Doorlock<br />

Group 1 (their assignment cannot be changed). A mechanical interlock<br />

ensures that scripts and processes cannot be started unless the safety hoods<br />

are closed and locked. Two additional doorlocks can be fitted to the pipetting<br />

instrument to support special hardware arrangements.<br />

Specify the required doorlock group assignment for doorlocks 3 and 4 by<br />

choosing the required group from the list. Choose none if the additional<br />

doorlocks are not fitted to your instrument.<br />

8.4.2.8 Arm Initialization<br />

Each time <strong>Freedom</strong> <strong>EVOware</strong> or an <strong>Freedom</strong> <strong>EVOware</strong> script or process is<br />

started, the status of the instrument arms (LiHa, RoMa and PnP) and optional<br />

devices (e.g. Te-MO and Te-VacS) is checked and they are initialized if necessary.<br />

In addition, a full initialization sequence is carried out when <strong>Freedom</strong> <strong>EVOware</strong> is<br />

started for the first time after powering up the instrument.<br />

When the instrument arms are initialized, a sequence of movements is carried out<br />

to determine the reference (zero) positions along the X, Y and Z axes. During the<br />

full initialization sequence, the arms need to move fully to the rear of the<br />

instrument to determine the zero position in the Y direction.<br />

You can also initialize the instrument manually by clicking the Initialize button in<br />

the toolbar (see 6.7 “<strong>Freedom</strong> <strong>EVOware</strong> Toolbar”, 6-18).<br />

A progress bar shows the progress of the initialization:<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 8 - 31