A - Appendix A<br />

Advanced Worklist Commands<br />

Tab. A-43 Advanced worklist command Wash-Multipip, parameters (cont.)<br />

Par. Name Type Range Explanation<br />

8 Washliquid string1 Wash liquid to use in step (container 1 or<br />

container 2)<br />

9 Flowthrough string1 1 - 100 WRC flow rate for step (%)<br />

10 Change string1 Wash liquid change for cleaner step<br />

(ignored for empty tips step: pad with 0)<br />

11 Preflush string1 0 - 1000 Preflush time for step, seconds<br />

12 Postflush string1 0 - 1000 Postflush time for step, seconds<br />

a) The minimum (cleanerSteps=1) is one Empty Tips step followed by one Cleaner step. This requires two<br />

sets of the parameters 2 - 12.<br />

Example with one Empty Tips step and two Cleaner steps:<br />

Wash-Multipip(2,22,9,"384 DMSO",5,1,1,1,40,0,0,0,22,9,"384<br />

DMSO",5,5,1,1,50,0,5,0,22,9,"384 DMSO",5,1,1,1,50,0,0,0);<br />

A.15.4.30 Te-MO 384 Wash Tip Head (Worklist: Wash-TeMO-384)<br />

The advanced worklist command which corresponds to the Te-MO 384 Wash Tip<br />

Head script command is “Wash_Multipip_384”. This advanced worklist command<br />

can only be used in <strong>Freedom</strong> <strong>EVOware</strong> <strong>Standard</strong>.<br />

The parameters of the advanced worklist command are shown in the following<br />

table. For more information on this command see 15.56.5 “Te-MO 384 Wash Tip<br />

Head Command”, 15-130.<br />

Tab. A-44 Advanced worklist command Wash_Multipip_384, parameters<br />

Par. Name Type Range Explanation<br />

1 flushTime floating point 0.1 - 50 Flush time in wash block<br />

2 flushTimeout floating point 0.1 Always specify 0.1<br />

3 washTime floating point 0.1 - 50 Wash time in wash block<br />

4 washTimeout floating point 30 Always specify 30<br />

5 washStationID integer 0 - grid position and site of the wash bloc,<br />

coded as follows: washStationID = grid *<br />

256 + site<br />

6 quickFlush integer 0 - 1 1 = Use quick flush<br />

7 plungerWash integer 1 = Tip washing with plunger moves<br />

8 washCyles string1 Wash cycles<br />

9 washVolume string1 1 - 100 Wash volume<br />

10 washSpeed string1 Wash speed<br />

12 postflush integer 0 - 1 1 = Flush tips after plunger moves<br />