15 - Script Commands<br />

Export Data Command<br />

the separate error codes. The Te-MO does not have a barcode scanner and does<br />

not generate barcode scanning errors.<br />

Tab. 15-7 Barcode scanning error codes in the report file output by the Export Data command<br />

Code<br />

(decimal)<br />

Code<br />

(hex) Caused by Explanation<br />

0 0 - No error.<br />

16 10 Instrument A barcode could not be read by the PosID or the<br />

PosID-3 scanner.<br />

32 20 Instrument The same labware was scanned again (at the same<br />

grid position or at a new worktable position of the<br />

same labware) and the scanned barcode did not<br />

match the previously scanned barcode. If this error<br />

occurs, <strong>Freedom</strong> <strong>EVOware</strong> updates the stored information<br />

for the labware with the newer barcode.<br />

64 40 Instrument An identical barcode was scanned at an unexpected<br />

grid position. Either (a) two labware items have identical<br />

barcodes or (b) the labware was moved to a different<br />

grid position manually. If this error occurs,<br />

<strong>Freedom</strong> <strong>EVOware</strong> assumes that (b) has taken<br />

place and updates the stored information for the labware<br />

with the new position.<br />

4096 1000 User The user entered the barcode manually after a scanning<br />

error.<br />

8192 2000 User The user chose Retry after a barcode scanning<br />

error.<br />

16384 4000 User The user chose Ignore after a barcode scanning<br />

error.<br />

The following table shows possible codes in the Error column. The value is bitcoded.<br />

If several errors occurred, the value shown is the bit-wise addition of the<br />

separate error codes.<br />

Tab. 15-8 Pipetting error codes in the report file output by the Export Data command<br />

Code<br />

(decimal)<br />

Code<br />

(hex) Caused by Explanation<br />

0 0 - No error.<br />

1 1 Instrument No liquid was detected.<br />

2 2 Instrument Not enough liquid was detected.<br />

4 4 Instrument Clot was detected.<br />

8 8 Instrument A dilutor error occurred during pipetting.<br />