9 - Specifying Your Instrument Setup<br />

Configuring Carriers<br />

and 10 and “Stack2” to sites 14 and 15 (see Tab. 8-9 “Te-MO 3x5 site<br />

numbers”, 8-104). Choose descriptive labels with no space characters. See<br />

also <strong>EVOware</strong> location in 8.5.28.4 “Te-Stack and Te-Stack Te-MO<br />

Configuration - Locations Tab”, 8-102.<br />

A group is also needed for the Te-MO if you are using Active Positioning<br />

Carriers (APCs) but you have not fitted them to all sites. Assign the group<br />

label “APC” to the APC sites and put the labware onto one of the APC sites. if<br />

<strong>Freedom</strong> <strong>EVOware</strong> decides to use another Te-MO site or slide for pipetting, it<br />

will only use the APC sites.<br />

The location groups you have created will be shown in dialog boxes which<br />

allow you to choose locations. The group name you have chosen is shown<br />

appended to the end of the carrier name, e.g. Te-MO_3x5_Stack1.<br />

You can also use a label in the Groups column to identify a single site.<br />

For example, you can assign a group label (e.g. “DITI-Waste”) to the waste<br />

site of a DITI carrier if you want <strong>Freedom</strong> <strong>EVOware</strong> Plus to use it for<br />

discarding racks with used DITIs. You must then specify this site for<br />

MoveLocation2 in the respective RoMa_ReplaceObject command (see<br />

9.10.3.1 “Editing Custom Pre-actions and Post-actions”, 9-74, “Usage<br />

Example”). If you don’t specify a specific waste site, <strong>Freedom</strong> <strong>EVOware</strong> Plus<br />

will return DITI racks with used DITIs to their base (labware source) location,<br />

e.g. the carousel.<br />

If you want to use group labels, you must assign labels to all sites of the<br />

carrier, not just one site. See e.g. Tab. 8-11 “Te-MO 3x5 group labels”, 8-<br />

104. For more examples, refer to Group labels for carrier sites in the index.<br />

The Groups function is not relevant for <strong>Freedom</strong> <strong>EVOware</strong> <strong>Standard</strong>.<br />

Allowed Labware on this carrier<br />

For each of the labware types which can be used with this labware carrier or<br />

device carrier, check the checkboxes alongside the allowed sites.<br />

9.5.1 Carrier Dimensions and X/Y Offsets<br />

The figure below shows how to define the various dimensions for a carrier. The<br />

example is for the standard Tecan carrier Microplate, 3 Pos., landscape. The<br />

carrier type which is shown has a nose which extends beyond the rear edge. The<br />

positioning pins (shown in grey) are clipped to the worktable and are partially<br />

hidden from view by the carrier.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 9 - 45