10 - Configuring Liquid Classes<br />

Te-MO 96 Liquid Classes<br />

Retract Speed<br />

Specify the required retract speed.<br />

10.5.3 Te-MO 96 Calibration Parameters<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses these parameters for the 96-tip Te-MO tip block or DITI<br />

pipetting head to ensure that the specified net volumes are dispensed accurately.<br />

The two curves in the dialog box show graphically the current settings for single<br />

pipetting mode and multi pipetting mode.<br />

Fig. 10-17 Te-MO 96 Calibrate parameters for the tip block and DITIs<br />

The parameters of this dialog box are as follows (some of the parameters must be<br />

specified separately for single pipetting mode and multi pipetting mode):<br />

Offset / Factor<br />

Specify an offset and/or calibration factor to be applied for single pipetting<br />

mode and multi pipetting mode.<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses the following expression to calculate the Te-MO<br />

plunger movement from the net volume specified in the pipetting script:<br />

Calibrated Te-MO plunger movement = (net volume * factor) + offset<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 10 - 37