18 - Error Handling<br />

List of Error Messages<br />

Tab. 18-1 List of error messages (cont.)<br />

Error ID Message Text Reason Corrective Measure<br />

EVO_SBC_000_002<br />

Generated ID could not be<br />

saved in DB<br />

Labware not found in the<br />

database or labware<br />

locked<br />

• Restart <strong>EVOware</strong><br />

• Restore backup<br />

• Reinstall <strong>EVOware</strong><br />

EVO_SBC_000_003<br />

Error accessing labware '' in<br />

the database.<br />

Labware not found in the<br />

database<br />

• Restart <strong>EVOware</strong><br />

• Restore backup<br />

• Reinstall <strong>EVOware</strong><br />

EVO_SBC_000_004 Execution was successful n.a. n.a.<br />

EVO_SBC_000_005 Error occured during execution n.a. (protocol message) n.a.<br />

Scheduler<br />

EVO_SCH_000_000 Halt pressed, system Interrupted! n.a. n.a.<br />

EVO_SCH_000_001 Module registered. n.a. n.a.<br />

EVO_SCH_000_002 Instrument is ready. n.a. n.a.<br />

EVO_SCH_000_003 Loading module n.a. n.a.<br />

EVO_SCH_000_004 GetInstrument: Index out of range. GetInstrument requested<br />

with wrong index<br />

EVO_SCH_000_005 GetInstrument: Bad index. GetInstrument requested<br />

with wrong index<br />

Contact your local service<br />

organization<br />

Contact your local service<br />

organization<br />

EVO_SCH_000_006<br />

LoadInstruments is currently not<br />

allowed.<br />

Inernal error<br />

Contact your local service<br />

organization<br />

EVO_SCH_000_007<br />

LoadAllInstruments is currently not<br />

allowed.<br />

Inernal error<br />

Contact your local service<br />

organization<br />

EVO_SCH_000_008<br />

UnloadInstruments is already running.<br />

Inernal error<br />

Contact your local service<br />

organization<br />

EVO_SCH_000_009<br />

UnloadInstruments can not be run at<br />

this time.<br />

Inernal error<br />

Contact your local service<br />

organization<br />

EVO_SCH_000_010 Initialization currently not allowed. Inernal error Contact your local service<br />

organization<br />

EVO_SCH_000_011<br />

Configuration file has been changed.<br />

Initialization aborted.<br />

Trac.ini has been modified<br />

Restore originial file or<br />

open configuration and<br />

accept changes<br />

EVO_SCH_000_012<br />

Error initializing instrument drivers.<br />

Instrument driver '' unreachable!:<br />

<br />

Communication to driver<br />

failed<br />

• Restart <strong>EVOware</strong><br />

• Reboot computer<br />

• Reinstall <strong>EVOware</strong><br />

EVO_SCH_000_013 Could not start scheduler Scheduler could not be<br />

connected or shut down<br />

Restart <strong>EVOware</strong><br />

EVO_SCH_000_014<br />

Error shutting down instruments.<br />

Instrument driver '' unreachable!:<br />

<br />

Communication to driver<br />

failed<br />

If error recurs contact<br />

your local service organization<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 18 - 63