15 - Script Commands<br />

Export Data Command<br />

Tab. 15-8 Pipetting error codes in the report file output by the Export Data command (cont.)<br />

Code<br />

(decimal)<br />

Code<br />

(hex) Caused by Explanation<br />

256 100 Instrument Liquid Arrival Check, liquid error: The dispensed volume<br />

was outside of the tolerance specified in the liquid<br />

class. This error is generated by the LAC<br />

(pooling balance) feature (see 15.26.4 “Pooling balance<br />

data in report files”, 15-56).<br />

512 200 Instrument Liquid Arrival Check, technical error: timeout or<br />

hardware malfunction such as balance overload.<br />

This error is generated by the LAC (pooling balance)<br />

feature.<br />

1,024 400 Instrument The PMP option detected a tip orifice occlusion (e.g.<br />

clot aspirated). This error also results in error code 4<br />

(clot detected).<br />

2,048 800 Instrument The PMP option detected an incorrect pipetting volume.<br />

65,536 10000 User The dilutor was initialized after a dilutor error.<br />

131,072 20000 User The dilutor was deactivated after a dilutor error.<br />

262,144 40000 User The user chose Retry after a clot-detection error.<br />

524,288 80000 User The user chose Ignore after a clot-detection error.<br />

1,048,576 100000 User The user chose Continue after a clot-detection error.<br />

2,097,152 200000 User The user chose Retry after a liquid-detection error.<br />

4,194,304 400000 User The user chose Aspirate air after a liquid-detection<br />

error.<br />

8,388,608 800000 User The user chose Move tips to Z-Max after a liquiddetection<br />

error.<br />

33,554,432 2000000 The PMP option detected an instrument error (e.g. a<br />

leakage from the air gap or a blocked pressure<br />

channel). a)<br />

67,108,864 4000000 The PMP option detected an unspecified error. If this<br />

occurs more than three times in a row for the same<br />

tip, this can also indicate an instrument error. a)<br />

a) In the case of an instrument error, you can use the Setup & Service software (PMP panel) to get more<br />

detailed information on the error. The Setup & Service software is provided by Tecan to service engineers.<br />

See 17.6.4 “PMP Error and Success Messages”, 17-22 for more information on PMP errors and error<br />

handling.<br />

Note: If a liquid from a container or tube which returns a barcode error is pipetted<br />

to another labware item, the error is inherited in the destination well as a pipetting<br />

error (not a barcode error). The value of the error code is unchanged (see<br />

Tab. 15-7, 15-54 for a list of the applicable error codes).<br />

Note: If “Pipette Nothing” was chosen after a liquid level detection error or clot<br />

error, the affected well(s) will not be reported any more in the output files.<br />