16 - Device Commands<br />

Commands for the Te-VacS<br />

16.35.10 Te-VacS - PositionPlateRear Command<br />

The PositionPlateRear command is only relevant if the Te-VacS is fitted with the<br />

Repositioner option. This option allows the contents of four subsequent 96-well<br />

SPE plates (e.g. TecPrep 96 microplates) to be extracted to one 384-well<br />

collection plate (e.g. a MALDI target). The collection plate is placed on a movable<br />

carrier.<br />

The PositionPlateRear command moves the collection plate for the rear vacuum<br />

block into one of four positions before you apply the vacuum. After extracting<br />

liquid from one of the 96-well microplates, it is replaced by another microplate and<br />

the collection plate is moved by 4.5 mm in the X and/or Y direction to the new<br />

position.<br />

The collection plate can be moved to the following positions:<br />

• left/rear<br />

• left/front<br />

• right/rear<br />

• right/front.<br />

Execution parameters<br />

You are prompted to specify the following execution parameters when you add<br />

this command to your script or process:<br />

Fig. 16-85 Te-VacS - PositionPlateRear command<br />

Position<br />

Choose the required position of the collection plate.<br />

To view or edit the execution parameters again later, right-click on the script<br />

command in the Script Editor and choose Edit in the context menu.<br />

16.35.11 Te-VacS - PositionPlateFront Command<br />

The PositionPlateFront command is only relevant if the Te-VacS is fitted with the<br />

Repositioner option. This option allows the contents of four subsequent 96-well<br />

SPE plates (e.g. TecPrep 96 microplates) to be extracted to one 384-well<br />

collection plate (e.g. a MALDI target). The collection plate is placed on a movable<br />

carrier.<br />

The PositionPlateFront command moves the collection plate for the front vacuum<br />

block into one of four positions before you apply the vacuum. After extracting<br />

liquid from one of the 96-well microplates, it is replaced by another plate and the<br />

collection plate is moved by 4.5 mm in the X and/or Y direction to the new<br />

position.<br />

The collection plate can be moved to the following positions:<br />

• left/rear<br />

• left/front<br />

• right/rear<br />