8 - Configuration Tool<br />

Device Settings<br />

ATTENTION<br />

Risk of the tips colliding with the waste and/or the washer.<br />

• Please check the coordinates of the waste and/or the washer before<br />

activating these options. Otherwise the tips might collide with the labware.<br />

DITI Handling panel<br />

Optimize positions when fetching DITIs<br />

If you check this checkbox, the following method is used to reduce the number<br />

of LiHa movements which are needed when getting DITIs (i.e. to speed up the<br />

Get DITI commands):<br />

If you choose all of the tips in the Get DITI command, they are picked up<br />

starting at the beginning of the DITI rack (for the 8-tip LiHa: positions 1 to 8).<br />

The next eight tips are then picked up at positions 9 to 16.<br />

If you do not choose all of the tips in the Get DITI command, they are picked<br />

up at the end of the DITI rack (starting at position 96).<br />

This feature ensures that the beginning of the DITI rack always has full rows<br />

of fresh DITIs which can be picked up in a single movement.<br />

This optimization feature is permanently activated for the <strong>Freedom</strong> EVO 75<br />

pipetting instrument with the 2-tip LiHa or 8 PLus 1 Access option (see<br />

A.7 “<strong>Freedom</strong> EVO 75 Pipetting Instrument - LiHa Configuration”, A-30).<br />

DITI fetch mode<br />

Specify the required DITI fetch mode as follows:<br />

Sequential: The DITIs are mounted one after the other sequentially..<br />

Even/Odd: All of the even-numbered DITIs are mounted simultaneously (2, 4,<br />

6, 8) followed by all of the odd-numbered DITIs simultaneously (1, 3, 5, 7).<br />

This option requires a reinforced DITI support bracket (hardware option).<br />

All parallel: All of the DITIs are mounted simultaneously. This option is only<br />

allowed for mounting up to four DITIs (maximum). In addition, this option<br />

requires a reinforced DITI support bracket (hardware option).<br />

The options Even/Odd and All Parallel reduce the time required to mount the<br />

DITIs (increased performance).<br />

Speed up Movements panel<br />

Use fast positioning in XZ<br />

When the LiHa moves across a single labware item in the X direction, for<br />

example for reagent distribution, the tips are normally moved upwards (Z<br />

direction), then across to the next column of wells and then down again.<br />

Check this checkbox to enable smart moves. In this case, the X and Z<br />

movements take place simultaneously and the tips follow a curved or<br />

parabolic path instead of a rectangular path. This reduces the time required<br />

for the movement.<br />