15 - Script Commands<br />

Te-MO Commands<br />

15.56.7 Te-MO Drop DITIs Command<br />

The Te-MO Drop DITIs command is used to return the disposable tips to a<br />

specified position on the Te-MO.<br />

Fig. 15-91 Te-MO Drop DITIs command<br />

The parameters of the Te-MO Drop DITIs command are as follows:<br />

Back to source<br />

This option returns the DITIs back to the Te-MO carrier and slide position from<br />

where they were fetched with the Te-MO Get DITIs command.<br />

Note: If you switch off the pipetting instrument with the DITIs still attached, it will<br />

forget the Te-MO carrier and slide position from where the DITIs were fetched.<br />

The option Back to source will then display an error message when the script is<br />

validated.<br />

Back to position<br />

Specify the position where the Te-MO should return the DITIs by clicking on<br />

the Te-MO DITI rack in the Worktable Editor. The type or label (name), the Te-<br />

MO grid position and the site of the DITI rack are then shown in the gray<br />

(protected) fields, see also Tab. 8-9 “Te-MO 3x5 site numbers”, 8-104. You<br />

must first put a Te-MO DITI rack in the Worktable Editor at the required<br />

position. You can only put Te-MO DITI racks on the middle slide of the Te-MO.<br />

Discard DITIs into waste<br />

Check this checkbox if you want to discard the DITIs after returning them to<br />

the specified position. After putting down the DITIs in the DITI rack, the rack<br />

will be pushed into the waste chute. This operation requires the DITI rack to<br />

be positioned on the front most site of the centre slide.<br />

15.56.7.1 Side movements when dropping Te-MO DITIs<br />

Under unusual environmental conditions (e.g. very dry ambient air), Te-MO DITIs<br />

may not drop reliably due to electrostatic charging of the plastic tips. In such<br />

cases, you can specify an optional side and/or vertical movement of the Te-MO<br />

pipetting head directly after releasing the tips.<br />

This is done by adding entries to the [TEMO] section of the configuration file<br />

evoware.opt using a text editor. evoware.opt is in the \Database<br />

directory.<br />

Add the following line to move by 10 instrument steps (10 x 0.1 mm) in the X<br />

direction:<br />

DITI_SIDE_STEPS_X=10<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 15 - 133