15 - Script Commands<br />

Wash Tips Command<br />

Use Waste<br />

Specify the worktable position for the waste you want to use by clicking on it in<br />

the Worktable Editor. You must first put a wash station with waste unit in the<br />

Worktable Editor at the required position. Alternatively, choose the required<br />

waste position from the pull down list. The list shows all waste units which are<br />

currently positioned in the Worktable Editor.<br />

The gray (protected) fields show the worktable coordinates of the selected<br />

waste.<br />

Use Cleaner<br />

Specify the worktable position for the cleaner you want to use by clicking on it<br />

in the Worktable Editor. You must first put a wash station with cleaner unit in<br />

the Worktable Editor at the required position. Alternatively, choose the<br />

required cleaner position from the pull down list. The list shows all cleaner<br />

units which are currently positioned in the Worktable Editor.<br />

The gray (protected) fields show the worktable coordinates of the selected<br />

cleaner.<br />

Choose a shallow cleaner if you only need to clean the ends of the tips.<br />

Choose a deep cleaner if there is a possibility of contamination along the shaft<br />

of the tip. The deep cleaner requires a larger volume of system liquid and<br />

cleaning takes somewhat longer.<br />

The wash cycle is skipped if you are flushing DITI adapters and Use Cleaner<br />

is ignored in this case.<br />

Use Fast-Wash Module<br />

If your instrument is fitted with FWO (Fast Wash Option) or MPO (Monitored<br />

Pump Option), check this checkbox to use the Fast Wash module instead of<br />

the dilutors to wash the tips. If you use the Fast Wash module, the volumes of<br />

liquid which are dispensed in the flushing and washing steps may differ<br />

significantly from the volumes you specify. This is due to accuracy limitations<br />

of the pump.<br />

Use Pinch Valve<br />

If your instrument is fitted with the Low Volume Option, check this checkbox to<br />

activate the Low Volume Option after washing the tips.<br />

Delay in Waste<br />

To reduce the pressure inside the tubing, you can instruct the instrument to<br />

keep the valves open somewhat longer after flushing before moving the LiHa<br />

away from the waste (range: 0 - 1000 ms).<br />

Delay in Cleaner<br />

To reduce the pressure inside the tubing, you can instruct the instrument to<br />

keep the valves open somewhat longer after washing before moving the LiHa<br />

away from the cleaner (range: 0 - 1000 ms). The wash cycle is skipped if you<br />

are flushing DITI adapters and Delay in Cleaner is ignored in this case.<br />

Retract with<br />

Specify the speed with which you want to retract the tips from the cleaner<br />

(range: 0 - 100 mm/s). If the speed is too high, droplets may be dragged out of<br />

the cleaner and stick to the outside of the tips.<br />

Airgap volume, Airgap speed<br />

These settings are used to specify a system trailing airgap (STAG) which will<br />

be aspirated after washing the tips. Specify the required size of the airgap in µl<br />

and the aspiration speed for the airgap in µl/s. The system trailing airgap is<br />

used to prevent contamination by ensuring that the system liquid cannot come<br />

into contact with the liquid which will be aspirated in a subsequent pipetting<br />

command.<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 15 - 9