15 - Script Commands<br />

Serial Dilution Wizard<br />

Select the waste site<br />

Choose the waste unit you want to use from the pull down list. The list shows<br />

all waste units which are currently positioned in the Worktable Editor. The<br />

worktable grid position is shown in brackets. You must first put a wash station<br />

with waste unit in the Worktable Editor.Alternatively, specify the worktable<br />

position for the waste by clicking on it in the Worktable Editor. If only one<br />

waste unit is present, it is chosen automatically.<br />

When using multi-pipetting mode, <strong>Freedom</strong> <strong>EVOware</strong> uses the chosen waste<br />

for discarding the excess volume before aspirating again from the same<br />

source. See also 15.5 “Periodic Wash Command”, 15-10.<br />

Select the cleaner site<br />

This option only appears if you have if you are using fixed tips. Choose the<br />

cleaner unit you want to use from the pull down list. The list shows all cleaner<br />

units which are currently positioned in the Worktable Editor. You must first put<br />

a wash station with cleaner unit in the Worktable Editor. If only one cleaner<br />

unit is present, it is chosen automatically. Alternatively, specify the worktable<br />

position for the cleaner by clicking on it in the Worktable Editor.<br />

Choose a shallow cleaner if you only need to clean the ends of the tips.<br />

Choose a deep cleaner if there is a possibility of contamination along the shaft<br />

of the tip. The deep cleaner requires a larger volume of system liquid and<br />

cleaning takes somewhat longer.<br />

Volumes in cleaner and in waste<br />

These options only appear if you are using fixed tips. Specify the volume of<br />

system liquid which should be used to wash the outside of the tips (volume in<br />

cleaner) and to flush the inside of the tips (volume in waste). Appropriate<br />

settings are chosen automatically. For more information on washing see<br />

15.4 “Wash Tips Command”, 15-8.<br />

Use Fast-Wash Module<br />

If your instrument is fitted with FWO (Fast Wash Option) or MPO (Monitored<br />

Pump Option), check this checkbox to use the Fast Wash module instead of<br />

the dilutors to wash the tips.<br />

Use Pinch Valve<br />

If your instrument is fitted with the Low Volume Option, check this checkbox to<br />

activate the Low Volume Option after washing the tips.<br />

Do you want to mix after dispensing<br />

Click the appropriate radio button. If you choose Yes, you must also specify<br />

how many times to mix and the volume to mix with.<br />