12.01.2014 Views

Software Manual Freedom EVOware Standard 2.3 Freedom ...

Software Manual Freedom EVOware Standard 2.3 Freedom ...

Software Manual Freedom EVOware Standard 2.3 Freedom ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

A - Appendix A<br />

Parameters in the Evoware.opt file<br />

A.3 Parameters in the Evoware.opt file<br />

The following tables list user-configurable parameters in six different sections of<br />

the evoware.opt file. evoware.opt is in the /Database directory.<br />

These settings should only be changed by specialists.<br />

Note: The parameters in evoware.opt are already configured with suitable values.<br />

The values should only be changed for special applications. Do not change the<br />

values of parameters which are not listed in the following tables.<br />

Add a new line to the respective section of the evoware.opt file if the parameter is<br />

missing.<br />

Default values: Each parameter string is shown below with the default value of<br />

the parameter (some parameters do not have a default value). The parameter<br />

string is not case sensitive.<br />

Note: <strong>Freedom</strong> <strong>EVOware</strong> must be re-started for the changes to take effect.<br />

Configuration files are protected with checksums. When you re-start <strong>Freedom</strong><br />

<strong>EVOware</strong> after editing this file, you will be warned that the checksum is invalid.<br />

Proceed as follows to repair the checksum:<br />

• Quit <strong>Freedom</strong> <strong>EVOware</strong> and start the Configuration Tool with Start > All<br />

Programs > Tecan > <strong>EVOware</strong> > Configure. You will be warned again that the<br />

checksum is invalid and you will be asked if you want to repair the checksum.<br />

Click Yes. Then Click Exit and re-start <strong>Freedom</strong> <strong>EVOware</strong>. Alternatively, repair<br />

the checksum using the Validate Configuration tool (see A.9.2 “Validate<br />

Configuration Tool”, A-36).<br />

Tab. A-2 Evoware.opt parameters, [OPTIONS] section<br />

Parameter string Description Range, units<br />

Lauda_Pump_Speed=3<br />

Csv_With_Semicolon=0<br />

Speed of pump for Lauda waterbath.<br />

1= Use “;” instead of “,” as the field<br />

separator in the CSV file created<br />

by the Export Data command. This<br />

parameter also controls the field<br />

separator for the Import Variable<br />

and Export Variable commands<br />

(see 15.42.2, 15-107).<br />

It has no influence on the scan.csv<br />

file exported by the PosID command,<br />

which always uses “;” as<br />

the field separator.<br />

1 to 5<br />

0 or 1<br />

Posid_Allocation_Timeout=60000 Time-out for collision avoidance. 0 to ∞, in ms<br />

LiHa_Allocation_Timeout=60000 Time-out for collision avoidance. 0 to ∞, in ms<br />

RoMa_Allocation_Timeout=60000 Time-out for collision avoidance. 0 to ∞, in ms<br />

PnP_Allocation_Timeout=60000 Time-out for collision avoidance. 0 to ∞, in ms<br />

Diluter_OFF=tipMask<br />

Turns the diluter off for the specified<br />

tip(s).<br />

0 to 255,<br />

bit-mapped<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> A - 17