10 - Configuring Liquid Classes<br />

Te-MO 384 Liquid Classes<br />

Fig. 10-20 Te-MO 384 Calibration parameters<br />

The parameters of this dialog box are as follows (some of the parameters must be<br />

specified separately for single pipetting mode and multi pipetting mode):<br />

Offset / Factor<br />

Specify an offset and/or calibration factor to be applied for single pipetting<br />

mode and multi pipetting mode.<br />

<strong>Freedom</strong> <strong>EVOware</strong> uses the following expression to calculate the Te-MO<br />

plunger movement from the net volume specified in the pipetting script:<br />

Calibrated Te-MO plunger movement = (net volume * factor) + offset<br />

10 - 40 <strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong>