16 - Device Commands<br />

Commands for the Safire Reader<br />

made with up to 1536-well microplates (max. labware height 24.5 mm). Switching<br />

between measurement techniques and labware formats is fully automated.<br />

The commands for this reader require the RdrOLE and XFluor4 software, which<br />

must be installed separately. <strong>Freedom</strong> <strong>EVOware</strong> can also control this reader<br />

using the Magellan software (see 16.16 “Commands for the Magellan option”,<br />

16-48).<br />

16.27.1 Safire Reader - Status Command<br />

The Status command interrogates the status of the reader. The device status is<br />

reported during process runtime in the Messages/Error window of the runtime<br />

controller (see 12.2.4 “Runtime Controller, Extended View (Running a Process)”,<br />

12-14).<br />

This command has no execution parameters or objects, pre-actions or postactions.<br />

16.27.2 Safire Reader - Open Command<br />

The Open command opens the door of the reader.<br />

This command has no execution parameters or objects, pre-actions or postactions.<br />

16.27.3 Safire Reader - Close Command<br />

The Close command closes the door of the reader.<br />

This command has no execution parameters or objects, pre-actions or postactions.<br />

16.27.4 Safire Reader - Measure Command<br />

The Measure command performs a measurement procedure using external<br />

software. For a detailed description of the parameters of this command please<br />

consult the documentation for RdrOLE and XFluor4.<br />

Execution parameters<br />

The RdrOLE reader interface software prompts you to specify the measurement<br />

parameters when you add this command to your script or process. Refer to the<br />

RdrOLE online help and manual for more information.<br />

Note: RdrOLE must be configured to connect to the chosen reader. In <strong>Freedom</strong><br />

<strong>EVOware</strong>’s offline mode, you must start RdrOLE and choose the reader manually.<br />

After making your entries in the Measurement Parameters dialog box and clicking<br />

Ok, the Output dialog box is shown where you must specify the name of the<br />

output file for the measurements. Alternatively, click Browse and choose an<br />

existing file. A default file name will be suggested automatically:<br />

<strong>Freedom</strong> <strong>EVOware</strong> <strong>Software</strong> <strong>Manual</strong>, 393172, en, V<strong>2.3</strong> 16 - 87