28.02.2013 Views

Handbook of Solvents - George Wypych - ChemTech - Ventech!

Handbook of Solvents - George Wypych - ChemTech - Ventech!

Handbook of Solvents - George Wypych - ChemTech - Ventech!

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

402 Seung Su Kim, Jae Chun Hyun<br />

Figure 7.3.17. Schematic <strong>of</strong> HADES. HADES contains several sections. [Adapted, by permission, Vinjamur and<br />

Cairncross, Presented at the AIChE national meeting, Dallas, Texas, November 1, 1999].<br />

Figure 7.3.18. Schematic <strong>of</strong> sample chamber <strong>of</strong> HADES. Dry nitrogen flows in from the port on the left side and<br />

the exhaust flows out through the port on the right to a FID. [Adapted, by permission, Winward and Cairncross, 9th<br />

International Coating Science and Technology Symposium, Delaware, USA, May 17 - 20, 1998, pp.343 - 346].<br />

solvent vapor concentration times the gas flow rate, and the solvent loss is found by integrating<br />

the evaporation rate.<br />

The HADES is consisted <strong>of</strong> several sections as shown in Figure 7.3.17; gas handling<br />

system, a sample chamber and several process measurements. 52,57 The nitrogen gas flows<br />

into the sample chamber as shown in Figure 7.3.18. The temperature and rate <strong>of</strong> gas flow are<br />

controlled and the gas temperatures before and after the coating sample tray are measured<br />

with thermocouples. The coating temperature is measured with thermocouple which is installed<br />

at the coating sample tray, and the solvent laden gas flows into the total hydrocarbon<br />

analyzer which is equipped with flame ionization detector (FID). The concentration <strong>of</strong> solvent<br />

at the exhaust gas is measured by total hydrocarbon analyzer, which is calibrated with<br />

known concentration <strong>of</strong> solvent vapor (via solvent bubblers).<br />

Figure 7.3.19 and Figure 7.3.20 show the examples <strong>of</strong> HADES running. 52 Figure<br />

7.3.19 shows the measured solvent concentration <strong>of</strong> the PVAC-toluene system, as shown in<br />

the figure the residual solvent decreases according to the rate <strong>of</strong> airflow. However, over the<br />

36 cm/s <strong>of</strong> airflow rate the drying rate wasn’t changed, above this airflow rate, the residual

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!