TRADITIONAL POSTER - ismrm
TRADITIONAL POSTER - ismrm
TRADITIONAL POSTER - ismrm
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Poster Sessions<br />
technique was found to decrease the Nyquist ghost by at least 75%, yielding images comparable to those obtained by using time-consuming, blade-specific<br />
reference scans.<br />
3101. Anisotropic Gradient Time Delay Correction for Oblique Radial Readouts Used in Ultrashort<br />
TE (UTE) Imaging<br />
Atsushi M. Takahashi 1<br />
1 Applied Science Laboratory, GE Healthcare, Menlo Park, CA, United States<br />
Gradient delays in MRI system are typically anisotropic and yield artifacts that are especially noticable in ramp sampled, center-out, radial k-space<br />
trajectories. We have developed a calibration procedure and a mathematical formulation for correcting artifacts from anisotropic gradient delays.<br />
3102. Correcting for Gradient Imperfections in Ultra-Short Echo Time Imaging<br />
Jeremy F. Magland 1 , Hamidreza Saligheh-Rad 1 , Felix W. Wehrli 1<br />
1 Department of Radiology, University of Pennsylvania Medical Center, Philadelphia, PA, United States<br />
Imperfections in readout gradients can cause scanner-specific problems in ultra-short echo time (UTE) imaging sequences. In addition to slight gradient<br />
delays, the shape of the readout gradient waveform may not be trapezoidal. Here we describe a simple technique for mapping the k-space trajectory of the<br />
initial readout ramp in a UTE pulse sequence. The method uses data from a short calibration scan in which two dimensions of spatial encoding is applied<br />
prior to readout. After correcting for B0 inhomogeneity, the method provides a very accurate measurement of the k-space trajectory during the ramp, which<br />
can be used as input to a gridding-based reconstruction algorithm.<br />
3103. Scaling in Readout Direction: A Vibration-Induced Distortion of Diffusion-Weighted Images and Its<br />
Retrospective Correction by Affine Registration<br />
Siawoosh Mohammadi 1 , Michael Deppe 1 , Harald E. Moller 2<br />
1 Department of Neurology, University of Muenster, Muenster, NRW, Germany; 2 Magnetic Resonance Unit, Max Planck Institute for<br />
Human Cognitive and Brain Sciences, Leipzig, Sachsen, Germany<br />
The strong lobes of the diffusion gradients cause different kinds of MR artifacts, like eddy-current (EC) and vibration effects. While EC effects could<br />
significantly be reduced using a twice-refocused spin-echo (TRSE) sequence for DTI acquisition, the vibration effects become more evident when the TRSE<br />
sequence is used. We showed that the vibration-induced motion leads to an affine scaling effect in x and y-direction that could be retrospectively corrected.<br />
While the y scaling is also subject to EC effects, the x scaling seems to correct solely vibration effects and might thus be usable for comparing vibration<br />
effects of different data sets.<br />
3104. Rapid Concomitant Field Correction for 2D Spiral Imaging<br />
Ajit Devaraj 1 , Payal Bhavsar 1 , James G. Pipe 1<br />
1 Keller Center for Imaging Innovation, Barrow Neurological Institute, Phoenix, AZ, United States<br />
Concomitant fields are a source of artifact for non-axial spiral images. The resulting artifacts are similar to Bo in-homogeneity blurring, hence challenging to<br />
account for. This work presents a rapid approach based on separable de-blur kernels. The efficacy of the proposed approach is demonstrated on both<br />
simulated and phantom sagittal images.<br />
3105. Efficient Off-Resonance Corrected Reconstruction of Rosette Trajectories by Deformed Interpolation<br />
Kernels<br />
Marco Reisert 1 , Jürgen Hennig 1 , Thimo Grotz 1 , Benjamin Zahneisen 1<br />
1 Medical Physics, University Hospital Freiburg, Freiburg, Baden-Wuerttemberg, Germany<br />
Using a 3D rosette trajectory and iterative, regularized reconstruction a 64 3 volume can be acquired in less than 30ms. Single shot trajectories suffer from<br />
off-resonance effects because of their long readout times. Common off-resonance correction methods approximate the phase map by a time segmentation to<br />
correct for these effects but slow down the reconstruction. We therefore have developed an off-resonance correction, which uses an approximation in space<br />
rather than in time by deforming k-space interpolation kernels leading to a speed up of a factor of 10 at comparable reconstruction quality.<br />
3106. MR Gradient Estimation Using a Linear Time Invariant Model<br />
Nii Okai Addy 1 , Holden H. Wu 1,2 , Dwight G. Nishimura 1<br />
1 Electrical Engineering, Stanford University, Stanford, CA, United States; 2 Cardiovascular Medicine, Stanford University, Stanford,<br />
CA, United States<br />
MR system imperfections limit the accuracy with which gradient waveforms of fast imaging trajectories such as spirals and 3D cones, are generated on the<br />
scanner. This mainly results in a delay of achieved k-space trajectories from the theoretical case. It is possible to measure the system delays for each axis and<br />
manually adjust the timing of the gradients to improve image reconstruction. However, a range of delay values can be observed on a single axis. This work<br />
models the gradient system with a linear time invariant model for accurate estimation of a range of gradient waveforms generated on the scanner.<br />
3107. Reference Coils Signal Combinations Removes Gradient Switching Artefacts in Physiological<br />
Recordings During MRI<br />
Roki Viidik 1,2 , Simon Bergstrand 3 , Tomas Karlsson 3 , Göran Starck 2,4<br />
1 Department of Signals and Systems, Chalmers University of Technology, Göteborg, Sweden; 2 Department of Medical Physics and<br />
Biomedical Engineering, Sahlgrenska University Hospital, Göteborg, Sweden; 3 Institute of Neuroscience and Physiology, Sahlgrenska<br />
University Hospital, Göteborg, Sweden; 4 Department for Radiation Physics, University of Gothenburg, Göteborg, Sweden<br />
Physiological registrations simultaneously with MR scanning usually require the removal of a huge gradient switching artefact from the weak physiological<br />
signal. We investigated a concept with pickup coils for simultaneous gradient switching registration for artefact removal. Adapted combinations of three<br />
reference signals recorded at the rear of the magnet could minimize the gradient artefact in all signal recordings at different positions in front of the magnet.<br />
The presented method works with any pulse sequence and any position and geometry of electrode leads loop.