WARNING<br />
ENSURE ALL MAIN POWER IS TURNED<br />
OFF BEFORE PROCEEDING WITH<br />
INSTALLATION. THIS UNIT IS<br />
PROVIDED WITH A MATING<br />
CONNECTOR FOR THE AC POWER<br />
SOCKET OR WITH A COMPATIBLE<br />
THREE WIRE GROUNDED CABLE<br />
WHICH MAY BE TERMINATED WITH A<br />
PLUG. ALWAYS ENSURE THE GROUND<br />
WIRE (GREEN OR GREEN AND<br />
YELLOW) OR GROUND PIN OF THE<br />
PLUG, IS CONNECTED TO A LOW<br />
IMPEDANCE SAFETY GROUND WITHIN<br />
THE AC POWER DISTRIBUTION<br />
SYSTEM. ALWAYS USE THE<br />
RECOMMENDED MATING CONNECTOR<br />
AND AN APPROVED THREE-WIRE<br />
CABLE TO CONNECT THIS UNIT TO<br />
THE MAIN AC.<br />
A. Main AC Socket - The mains voltage is<br />
connected via an appliance plug.<br />
WARNING<br />
THE APPLIANCE PLUG MUST BE<br />
CONNECTED WITH THE CORRECT<br />
POLARITY, I.E. THE CONNECTIONS L<br />
AND N MUST NOT BE INTERCHANGED.<br />
THE CORRECT CONNECTIONS ARE<br />
SHOWN IN FIGURE 2-6.<br />
THE CONNECTION TO THE BUILDING<br />
INSTALLATION MUST BE EITHER FIXED<br />
OR VIA A PLUG WHERE IT IS<br />
IMPOSSIBLE<br />
POLARITY.<br />
TO REVERSE THE<br />
Siemens AG C79000-G7374-C211-01<br />
<strong>SIREC</strong> D Manual<br />
104 022M A<br />
Installation and Wiring<br />
Figure 2-6 Application Plug<br />
Table 2-1 AC Wiring Color Codes<br />
COUNTRY GROUND Line<br />
(HOT)<br />
Neutral<br />
(RET)<br />
USA Green Black White<br />
EEC Green/<br />
Yellow<br />
Brown Blue<br />
B. Internal Fuse - This unit is equipped with an<br />
internal AC mains Fuse. If it should blow, it generally<br />
indicates a serious problem with the recorder<br />
and is not operator serviceable. The rear panel<br />
needs to be removed to replace the fuse which is<br />
on the power supply module. The fuse is a fast<br />
blow 5 x 20mm type rated at 2.5 A 250 VAC (~).<br />
2.3.3 Signal Input Wiring. Refer to the<br />
following paragraph.<br />
WARNING<br />
TO PREVENT THE POSSIBILITY OF<br />
ELECTRICAL SHOCK, USE EXTREME<br />
CAUTION WHEN WIRING SIGNAL INPUT<br />
CONNECTIONS. HAZARDOUS POTEN-<br />
TIALS MAY EXIST ON SIGNAL INPUT<br />
TERMINALS WHICH ARE FLOATING<br />
WITH RESPECT TO CASE GROUND.<br />
THESE HAZARDOUS POTENTIALS MAY<br />
BE ON THE REAR TERMINAL PANEL OF<br />
YOUR INSTRUMENT. ANY VOLTAGE<br />
POTENTIAL AT THE SIGNAL SOURCE<br />
WILL EXIST ON THE INSTRUMENT’S<br />
RESPECTIVE SIGNAL INPUT TERMINAL.<br />
E.G. POWER GENERATOR STATOR<br />
WINDING TEMPERATURE MONITORING<br />
THERMOCOUPLES. ENSURE THE<br />
POWER IS OFF BEFORE CONNECTING<br />
SIGNAL INPUTS TO THE UNIT.<br />
2-5