Creating an Instrument Control Method 101<br />

Lamp OFF at the end of run<br />

The option is disabled for the K-2700 detector. Check the Lamp OFF box<br />

to switch the detector lamp off after a measurement is finished. This<br />

option may be used for the shutdown methods.<br />

You should switch the lamp on from the direct control window<br />

before you start the next working session. After power up, allow 30<br />

minutes for the lamp to stabilize before running samples that may<br />

require sensitive measurements.<br />

<strong>Advanced</strong><br />

Pressing the <strong>Advanced</strong> button opens the window for reference<br />

correction:<br />

Fig. 149 reference correction window<br />

Referring to a wavelength with practically zero absorption the drift due for<br />

instance to a too short heating time of the lamp can be eliminated.<br />

Reference Correction<br />

According to the configuration up to 4 wavelengths can be monitored<br />

from the 3D data stream. Select the checkbox to activate the associated<br />

channel.<br />

Sample<br />

The sample wavelength is defined in the Acquisition tab.<br />

Bw<br />

The bandwidth for the sample channels are also defined in the<br />

Acquisition tab.<br />

Ref<br />

Reference is the wavelength at which a reference absorbance is<br />

measured as a function of time. The reference wavelength<br />

compensates for fluctuations caused by changes in baseline<br />

absorbance, for example due to changes in solvent composition<br />

during gradient elution. The selected wavelength must be higher than<br />

the sample wavelength and be inside of the scan range for this<br />

detector.<br />

Bw<br />

Enter the bandwidth of the reference wavelength. Limits are 1 to 32<br />

nm in 1 nm steps.<br />

Scan, nm [Range xxx – yyy nm]<br />

The displayed range depending on the detector type and the installed<br />

lamps will only be updated, if you open that window online.