Setup and Control of Knauer HPLC Systems 77<br />

Type<br />

Select the AS type 460 or 465 respectively 560 or 565. In the 300 series<br />

only 360 is possible.<br />

Serial Port<br />

Select the number for the communication port on your PC where the<br />

autosampler is connected.<br />

Port settings<br />

In the dialog box for the port settings only the baud rate will be active for<br />

selection of 4800 0r 9600.<br />

Fig. 107 Port settings<br />

Fig. 108<br />

Vial number<br />

For the AS 360, 560, and 565 a number in the range from 1 through 65<br />

can be selected. In a method only the number of vials entered here are<br />

available. A wrong number will lead to an error message:<br />

Vial type<br />

Select the volume of the used vials from the drop-down list. The choices<br />

are 1.8 ml, 0.5 ml, and 0.3 ml.<br />

Syringe Volume<br />

Select the volume of the installed syringe from the drop-down list. The<br />

choices are 250 µl, 500 µl, 1000 µl, and 2500 µl.<br />

Loop Volume<br />

Select the volume of the installed loop from the drop-down list. The<br />

choices are 20 µl, 100 µl, 500 µl, and 1000 µl.